Performance characteristics of the turbidimetric ACL-TOP CTS 300 coagulation analyzer in dogs and cats
Abstract
Objective: To evaluate the technical performance of the turbidimetric ACL-TOP CTS 300 coagulation analyzer (IL) in dogs and cats and to create reference intervals for standard and novel parameters.
Design: Coagulation testing results from dogs and cats generated by the IL were prospectively compared with another mechanical clot detection system. Precision was documented and reference intervals were created for prothrombin time, activated partial thromboplastin time, and Clauss fibrinogen and D-dimer values, as well as for the quantitative clot curve parameters (high and low amplitude, delta, and first and second derivative curves). Clot curve signatures containing common artifacts due to lipemia, hemolysis, or preactivation were demonstrated.
Animals: Residual frozen plasma from 20 dogs and 10 cats was used for method comparison; prospectively recruited healthy dogs (n = 48) and cats (n = 45) were used for reference interval and precision studies.
Measurements and main results: Bland-Altman analysis showed a proportional positive bias in the IL compared with mechanical clot detection in dogs and a suggestion of a similar pattern in cats. Precision was good and met manufacturer's recommendations for all assays. Reference intervals are reported. Clot curve artifacts were similar in animals to those reported in people.
Conclusions: The turbidimetric system had a slight high proportional positive bias compared with mechanical clot detection. Thus, new reference intervals were generated including for novel parameters generated by clot curve analysis. Some preanalytical errors can be identified by inspection of clot curves. This robust novel technology compares favorably with mechanical endpoint detection methods and can be used in dogs and cats.
Keywords: D-dimer; coagulopathy; fibrinogen; hemostatic testing; partial thromboplastin time; prothrombin time.
© Veterinary Emergency and Critical Care Society 2018.
Similar articles
-
Retrospective comparison of standard clotting tests and novel clot waveform parameters in dogs using the turbidimetric ACL-TOP CTS 300 coagulation analyzer.J Vet Diagn Invest. 2018 Nov;30(6):868-877. doi: 10.1177/1040638718797386. Epub 2018 Sep 11. J Vet Diagn Invest. 2018. PMID: 30204058 Free PMC article.
-
Retrospective characterization of canine coagulopathies using the turbidimetric ACL-TOP 300 analyzer (2014-2015): Forty-seven dogs.J Vet Emerg Crit Care (San Antonio). 2018 Nov;28(6):551-558. doi: 10.1111/vec.12775. Epub 2018 Oct 15. J Vet Emerg Crit Care (San Antonio). 2018. PMID: 30320489
-
Reference values for rotational thromboelastometry (ROTEM) in clinically healthy cats.J Vet Emerg Crit Care (San Antonio). 2017 Mar;27(2):185-192. doi: 10.1111/vec.12584. Epub 2017 Feb 6. J Vet Emerg Crit Care (San Antonio). 2017. PMID: 28166379
-
Viscoelastic coagulation testing: technology, applications, and limitations.Vet Clin Pathol. 2011 Jun;40(2):140-53. doi: 10.1111/j.1939-165X.2011.00302.x. Epub 2011 Mar 29. Vet Clin Pathol. 2011. PMID: 21446994 Review.
-
Specific assays of hemostasis proteins: fibrinogen.Ric Clin Lab. 1990 Apr-Jun;20(2):167-76. doi: 10.1007/BF02877563. Ric Clin Lab. 1990. PMID: 2196664 Review.
Cited by
-
Biomarkers of Coagulation and Inflammation in Dogs after Randomized Administration of 6% Hydroxyethyl Starch 130/0.4 or Hartmann's Solution.Animals (Basel). 2022 Oct 6;12(19):2691. doi: 10.3390/ani12192691. Animals (Basel). 2022. PMID: 36230433 Free PMC article.
-
The hemostatic profile of cold-stored whole blood from non-greyhound and greyhound dogs over 42 days.Front Vet Sci. 2023 Mar 2;10:1135880. doi: 10.3389/fvets.2023.1135880. eCollection 2023. Front Vet Sci. 2023. PMID: 36937016 Free PMC article.
-
Retrospective comparison of standard clotting tests and novel clot waveform parameters in dogs using the turbidimetric ACL-TOP CTS 300 coagulation analyzer.J Vet Diagn Invest. 2018 Nov;30(6):868-877. doi: 10.1177/1040638718797386. Epub 2018 Sep 11. J Vet Diagn Invest. 2018. PMID: 30204058 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous