Commercial Slit Lamp Anterior Segment Photography versus Digital Compact Camera Mounted on a Standard Slit Lamp with an Adapter
- PMID: 29909703
- DOI: 10.1080/02713683.2018.1490435
Commercial Slit Lamp Anterior Segment Photography versus Digital Compact Camera Mounted on a Standard Slit Lamp with an Adapter
Abstract
Purpose: To compare slit lamp mounted anterior segment cameras (SLCs) versus digital compact camera (DCC) with slit-lamp adaptor when used by an inexperienced technician.
Methods: In this cross sectional study, where posterior capsule opacification (PCO) was used as a comparator, patients were consented for one photograph with SLC and two with DCC (DCC1 and DCC2), with a slit lamp adaptor. An inexperienced clinic technician, who took all the photographs and masked the images, recruited one eye of each patient. Images were graded for PCO using EPCO 2000 software by two independent masked graders. Repeatability between DCC1 and DCC2, and limits-of-agreement between SLC and DCC1 mounted on slit-lamp with an adaptor were assessed. Coefficient-of-repeatability and Bland-Altmann plots were analyzed.
Results: Seventy-two patients (eyes) were recruited in the study. First 9 patients (eyes) were excluded due to unsatisfactory image quality from both the systems. Mean evaluation of posterior capsule opacification (EPCO) score for SLC was 2.28 (95% CI: 2.09-2.45), for DCC1 was 2.28 (95% CI: 2.11-2.45), and for the DCC2 was 2.11 (95% CI: 2.11-2.45). There was no significant difference in EPCO scores between SLC vs. DCC1 (p = 0.98) and between DCC1 and DCC2 (p = 0.97). Coefficient of repeatability between DCC images was 0.42, and the coefficient of repeatability between DCC and SLC was 0.58.
Conclusions: DCC on slit lamp with an adaptor is comparable to a SLC. There is an initial learning curve, which is similar for both for an inexperienced person. This opens up the possibility for low cost anterior segment imaging in the clinical, research, and teaching settings.
Keywords: Anterior segment photography; compact digital camera; photography.
Similar articles
-
Telemedical diagnosis of anterior segment eye diseases: validation of digital slit-lamp still images.Eye (Lond). 2009 Mar;23(3):652-60. doi: 10.1038/eye.2008.11. Epub 2008 Feb 15. Eye (Lond). 2009. PMID: 18274557
-
Stereoscopic three-dimensional (3D) slit-lamp photography using a compact 3D digital camera.Indian J Ophthalmol. 2021 May;69(5):1303-1305. doi: 10.4103/ijo.IJO_2037_20. Indian J Ophthalmol. 2021. PMID: 33913883 Free PMC article.
-
Assessing the subjective quality of smartphone anterior segment photography: a non-inferiority study.Int Ophthalmol. 2023 Feb;43(2):403-410. doi: 10.1007/s10792-022-02437-9. Epub 2022 Aug 26. Int Ophthalmol. 2023. PMID: 36018419 Free PMC article.
-
Anterior Segment Imaging Devices in Ophthalmic Telemedicine.Semin Ophthalmol. 2021 May 19;36(4):149-156. doi: 10.1080/08820538.2021.1887899. Epub 2021 Mar 3. Semin Ophthalmol. 2021. PMID: 33656960 Review.
-
On focusing the slit-lamp: Part I. An inaccurate ocular setting--what is there to lose?Surv Ophthalmol. 1998 Jan-Feb;42(4):351-4. doi: 10.1016/s0039-6257(97)00125-2. Surv Ophthalmol. 1998. PMID: 9493275 Review.
Cited by
-
Creating a secure clinical 'Bring Your Own Device' BYOD photography service to document and monitor suspicious lesions in the lid oncology clinic.Eye (Lond). 2023 Mar;37(4):744-750. doi: 10.1038/s41433-022-02049-8. Epub 2022 Apr 4. Eye (Lond). 2023. PMID: 35379923 Free PMC article.
-
Medical Students' Perspectives Regarding the Use of a Slit-Lamp Smartphone Adapter for Clinical Slit-Lamp Photography.Cureus. 2024 Apr 10;16(4):e57986. doi: 10.7759/cureus.57986. eCollection 2024 Apr. Cureus. 2024. PMID: 38606028 Free PMC article.
-
A novel approach to anterior segment imaging with smartphones in the COVID-19 era.Indian J Ophthalmol. 2021 May;69(5):1257-1262. doi: 10.4103/ijo.IJO_3707_20. Indian J Ophthalmol. 2021. PMID: 33913872 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources