Abstract
Objective: To assess the quality of mean uterine artery (UtA) pulsatility index (PI) measurement in a first-trimester pre-eclampsia screening program.
Methods: Consecutive women with a singleton pregnancy attending first-trimester screening for fetal chromosomal abnormalities also had combined screening for pre-eclampsia based on the Fetal Medicine Foundation (FMF) algorithm, at a large practice in Sydney, Australia, from May 2014 to February 2017. Distributions of mean UtA-PI multiples of the median (MoM) on a logarithmic scale were plotted in relation to the normal median with 95% CI for each operator and for each month. Central tendency and dispersion and cumulative sum charts were produced. Mean UtA-PI MoM values between 0.95 and 1.05 were considered ideal and those between 0.90 and 1.10 were considered acceptable. The screen-positive rates for preterm pre-eclampsia in different groups of sonographers according to their mean log10 UtA-PI MoM were calculated and compared using the chi-square test.
Results: A total of 21 010 women attended for first-trimester ultrasound and had screening for pre-eclampsia. The overall median UtA-PI MoM was 1.042 (interquartile range (IQR), 0.85-1.26). Of 46 sonographers, 42 (91.3%) performed more than 50 examinations and, of those, 41 (97.6%) measured UtA-PI within the acceptable range. Sonographers measuring UtA-PI MoM on average below 0.95 and those measuring it above 1.05 had, respectively, lower and higher screen-positive rates when compared with those with measurements within the 0.95-1.05 UtA-PI MoM interval (7.2% and 13.2% vs 11.2%, respectively, P < 0.001).
Conclusion: UtA Doppler is measured well among trained operators when following an established protocol. While slight variations are expected, systematic error in this measurement impacts on the screen-positive rate. Therefore, a quality control process should be in place and retraining of staff may be required. Copyright © 2018 ISUOG. Published by John Wiley & Sons Ltd.
Keywords: first trimester; pre-eclampsia; quality assessment; screening; uterine artery Doppler.
Copyright © 2018 ISUOG. Published by John Wiley & Sons Ltd.
Similar articles
-
Transverse technique: complementary approach to measurement of first-trimester uterine artery Doppler.Ultrasound Obstet Gynecol. 2018 Nov;52(5):639-647. doi: 10.1002/uog.18917. Ultrasound Obstet Gynecol. 2018. PMID: 28976627
-
Assessing quality standards in measurement of uterine artery pulsatility index at 11 to 13 + 6 weeks' gestation.Ultrasound Obstet Gynecol. 2015 Sep;46(3):299-305. doi: 10.1002/uog.14732. Ultrasound Obstet Gynecol. 2015. PMID: 25412757 Clinical Trial.
-
Screening for pre-eclampsia at 11-13 weeks' gestation: use of pregnancy-associated plasma protein-A, placental growth factor or both.Ultrasound Obstet Gynecol. 2020 Sep;56(3):400-407. doi: 10.1002/uog.22093. Epub 2020 Aug 5. Ultrasound Obstet Gynecol. 2020. PMID: 32441401
-
First trimester preeclampsia screening and prediction.Am J Obstet Gynecol. 2022 Feb;226(2S):S1071-S1097.e2. doi: 10.1016/j.ajog.2020.07.020. Epub 2020 Jul 16. Am J Obstet Gynecol. 2022. PMID: 32682859 Review.
-
First-trimester uterine artery Doppler and adverse pregnancy outcome: a meta-analysis involving 55,974 women.Ultrasound Obstet Gynecol. 2014 May;43(5):500-7. doi: 10.1002/uog.13275. Epub 2014 Apr 4. Ultrasound Obstet Gynecol. 2014. PMID: 24339044 Review.
Cited by
-
Notching and Pulsatility Index of the Uterine Arteries and Preeclampsia in Twin Pregnancies.J Clin Med. 2020 Aug 15;9(8):2653. doi: 10.3390/jcm9082653. J Clin Med. 2020. PMID: 32824222 Free PMC article.
-
Prevention of Pregnancy Complications Using a Multimodal Lifestyle, Screening, and Medical Model.J Clin Med. 2024 Jul 25;13(15):4344. doi: 10.3390/jcm13154344. J Clin Med. 2024. PMID: 39124610 Free PMC article. Review.
-
Role of Uterine Artery Doppler Ultrasound in Predicting Pre-Eclampsia in High-Risk Women.Cureus. 2021 Jul 9;13(7):e16276. doi: 10.7759/cureus.16276. eCollection 2021 Jul. Cureus. 2021. PMID: 34377610 Free PMC article.
-
Therapeutic Value of Low-Dose Acetylsalicylic Acid for the Prevention of Preeclampsia in High-Risk Bulgarian Women.Cureus. 2024 Aug 6;16(8):e66298. doi: 10.7759/cureus.66298. eCollection 2024 Aug. Cureus. 2024. PMID: 39113818 Free PMC article.
-
Predictive ability of serum advanced glycation end products at 11 to 13 weeks of gestation for early-onset preeclampsia.AJOG Glob Rep. 2022 Feb 26;2(2):100052. doi: 10.1016/j.xagr.2022.100052. eCollection 2022 May. AJOG Glob Rep. 2022. PMID: 36275494 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Research Materials
Miscellaneous