Reliability of Tidal Volume in Average Volume Assured Pressure Support Mode
- PMID: 29921607
- DOI: 10.4187/respcare.05917
Reliability of Tidal Volume in Average Volume Assured Pressure Support Mode
Abstract
Background: Remote monitoring is increasingly used in patients who receive home mechanical ventilation. The average volume assured pressure support mode is a target volume pressure preset mode that delivers a given tidal volume (VT) within a range of controlled inspiratory pressures. In a mode such as this, it is important to verify that the VT value retrieved from the ventilator SD card is accurate.
Methods: A lung model was set with C (Compliance) 0.075 L/cm H2O and RI (Inspiratory resistance)-RE (Expiratory resistance) 15-25 cm H2O/L/s (model 1) or with C 0.050 L/cm H2O and RI 6 cm H2O/L/s (model 2) and 6 cm H2O effort. Three home-care ventilators (A40, PrismaST30, and Vivo40) were set to average volume assured pressure support mode with 0.3 and 0.6 L VT each at PEEP 5 and 10 cm H2O, and were connected to the lung model with and without nonintentional leak. The reference airway pressure and flow were measured by a data logger. VT was expressed in body temperature and pressure saturated. We assessed the difference in VT between the ventilator SD card and a data logger relative to set VT and factors associated with its magnitude.
Results: For A40, PrismaST30, and Vivo40, the adjusted mean VT differences between the ventilator SD card and the data logger were -0.053 L (95% CI -0.067 to -0.039 L) (P < .001), -0.002 L (95%CI -0.022 to 0.019 L) (P = .86), and -0.067 L (95% CI -0.007 to 0.127 L) (P = .03), respectively. The partial Spearman correlation coefficients between the ventilator SD card and a data logger were 0.89 (P < .001), 0.59 (P < .001), and 0.78 (P < .001), respectively to the ventilators. The relative variations in measured VT from the set VT were 16.0, -12.0, and 6.7% for the ventilator SD card, and were -2.5, -7.5, and -27.2% for the data logger, respectively. The discrepancy in ventilator between SD card and data logger were influenced by PEEP for the PrismaST30 ventilator, nonintentional leak for the Vivo40 ventilator and PEEP, nonintentional leak, and underlying disease, the effect of each depending on the levels of the other factors, for the A40 ventilator.
Conclusions: In the 3 home-care ventilators, the ventilator SD card underestimated VT. Factors involved in this difference differed among the ventilators.
Keywords: average volume assured pressure support mode; bi-level; home mechanical ventilation; noninvasive ventilation.
Copyright © 2018 by Daedalus Enterprises.
Conflict of interest statement
The authors have disclosed no conflicts of interest.
Similar articles
-
Higher Fresh Gas Flow Rates Decrease Tidal Volume During Pressure Control Ventilation.Anesth Analg. 2017 May;124(5):1506-1511. doi: 10.1213/ANE.0000000000001944. Anesth Analg. 2017. PMID: 28333703 Free PMC article.
-
[Effects of aerosol inhalation on respiratory mechanical parameters under different ventilation patterns and ventilator parameters].Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2018 Nov;30(11):1036-1040. doi: 10.3760/cma.j.issn.2095-4352.2018.011.005. Zhonghua Wei Zhong Bing Ji Jiu Yi Xue. 2018. PMID: 30541641 Clinical Trial. Chinese.
-
Effect of inspiratory time on tidal volume delivery in anesthesia and intensive care unit ventilators operating in pressure control mode.J Clin Anesth. 2005 Feb;17(1):8-15. doi: 10.1016/j.jclinane.2004.02.005. J Clin Anesth. 2005. PMID: 15721723
-
The basics of respiratory mechanics: ventilator-derived parameters.Ann Transl Med. 2018 Oct;6(19):376. doi: 10.21037/atm.2018.06.06. Ann Transl Med. 2018. PMID: 30460250 Free PMC article. Review.
-
Noninvasive Ventilator Devices and Modes.Sleep Med Clin. 2020 Dec;15(4):545-555. doi: 10.1016/j.jsmc.2020.08.005. Epub 2020 Sep 26. Sleep Med Clin. 2020. PMID: 33131664 Review.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources