Abstract
Purpose: To investigate the impact of iterative metal artifact reduction (iMAR) on artifacts related to neurosurgical clips or endovascular coils when combined to filtered back projection (FBP) or advanced modelled iterative reconstruction (ADMIRE).
Material and methods: In this study 21 unenhanced brain computed tomography (CT) examinations were reconstructed with FBP and level 2 of ADMIRE, both techniques with and without iMAR algorithm, resulting in 4 series per acquisition. Subjective assessment of artifact reduction was performed as a double-blinded evaluation with a 5-point-scale. Objective analysis was performed by comparing central tendencies and distributions of voxel densities. The central tendency was assessed as the mean voxel density in Hounsfield units. The distribution was assessed by evaluating the shape and asymmetry of the histograms of voxels densities with measures of kurtosis and skewness, respectively.
Results: Inter-reader agreement was excellent (>0.8). FBP and ADMIRE without iMAR were scored 4 and with iMAR 5. Unusual artifacts were noted in all of the series reconstructed with iMAR, especially when combined with ADMIRE. Kurtosis revealed statistical differences for all reconstruction techniques (p ≤ 0.0007) except for the association of FBP with iMAR (p = 0.2211) for the coiling population and skewness demonstrated no statistical difference in any population (p ≥ 0.0558), confirming the subjective analysis results, except for the ADMIRE algorithm with or without iMAR (p ≤ 0.0342) in the coiling population.
Conclusion: iMAR led to the reduction in artifacts due to intracranial metallic devices. However, it created a new artifact in the form of a halo of photon-starvation, especially when combined with ADMIRE. The combination of FBP and iMAR seems more suitable, combining the beneficial metal artifact reduction without the emergence of a halo of photon starvation just around the point of interest.
Keywords: Artifacts; Computed tomography (CT); Computer-assisted algorithms; Intracranial aneurysm.
Similar articles
-
Metallic artifact reduction by evaluation of the additional value of iterative reconstruction algorithms in hip prosthesis computed tomography imaging.Medicine (Baltimore). 2019 Feb;98(6):e14341. doi: 10.1097/MD.0000000000014341. Medicine (Baltimore). 2019. PMID: 30732160 Free PMC article.
-
Iterative algorithms for metal artifact reduction in children with orthopedic prostheses: preliminary results.Pediatr Radiol. 2018 Dec;48(13):1884-1890. doi: 10.1007/s00247-018-4217-6. Epub 2018 Jul 28. Pediatr Radiol. 2018. PMID: 30056564
-
Efficiency of Iterative Metal Artifact Reduction Algorithm (iMAR) Applied to Brain Volume Perfusion CT in the Follow-up of Patients after Coiling or Clipping of Ruptured Brain Aneurysms.Sci Rep. 2019 Dec 19;9(1):19423. doi: 10.1038/s41598-019-55792-6. Sci Rep. 2019. PMID: 31857627 Free PMC article.
-
Artifact Reduction in Interventional Devices Using Virtual Monoenergetic Images and Iterative Metal Artifact Reduction on Photon-Counting Detector CT.Invest Radiol. 2025 Jul 1;60(7):463-469. doi: 10.1097/RLI.0000000000001149. Epub 2025 Jan 3. Invest Radiol. 2025. PMID: 39745893
-
[Iterative algorithms for artifact reduction in computed tomography].Radiologe. 2018 Mar;58(3):202-210. doi: 10.1007/s00117-017-0327-7. Radiologe. 2018. PMID: 29264618 Review. German.
Cited by
-
Metallic artifact reduction by evaluation of the additional value of iterative reconstruction algorithms in hip prosthesis computed tomography imaging.Medicine (Baltimore). 2019 Feb;98(6):e14341. doi: 10.1097/MD.0000000000014341. Medicine (Baltimore). 2019. PMID: 30732160 Free PMC article.
-
Photon-counting CT-angiography in comparison to digital subtraction angiography for assessing intracranial aneurysms after coiling or clipping.Neuroradiology. 2025 Jun 2. doi: 10.1007/s00234-025-03650-w. Online ahead of print. Neuroradiology. 2025. PMID: 40455108
-
Virtual monoenergetic images and post-processing algorithms effectively reduce CT artifacts from intracranial aneurysm treatment.Sci Rep. 2020 Apr 20;10(1):6629. doi: 10.1038/s41598-020-63574-8. Sci Rep. 2020. PMID: 32313094 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical
Miscellaneous