Value of high-resolution mapping in optimizing cryoballoon ablation of atrial fibrillation
- PMID: 29929934
- DOI: 10.1016/j.ijcard.2018.05.135
Value of high-resolution mapping in optimizing cryoballoon ablation of atrial fibrillation
Abstract
Background: Unrecognized incomplete pulmonary vein isolation (PVI), as opposed to post-PVI pulmonary vein reconnection, may be responsible for clinical recurrences of atrial fibrillation (AF). To date, no data are available on the use of high-resolution mapping (HRM) during cryoballoon (CB) ablation for AF as the index procedure. The aims of this study were: - to assess the value of using a HRM system during CB ablation procedures in terms of ability in acutely detecting incomplete CB lesions; - to compare the 8-pole circular mapping catheter (CMC, Achieve) and the 64-pole mini-basket catheter (Orion) with respect to pulmonary vein (PV) signals detection at baseline and after CB ablation; - to characterize the extension of the lesion produced by CB ablation by means of high-density voltage mapping.
Methods: Consecutive patients with drug-resistant paroxysmal or early-persistent AF undergoing CB ablation as the index procedure, assisted by a HRM system, were retrospectively included in this study.
Results: A total of 33 patients (25 males; mean age: 59 ± 18 years, 28 paroxysmal AF) were included. At baseline, CMC catheter revealed PV activity in 102 PVs (77%), while the Orion documented PV signals in all veins (100%). Failure of complete CB-PVI was more frequently revealed by atrial re-mapping with the Orion as compared to the Achieve catheter (24% vs 0%, p < 0.05). A repeat ablation was performed in 8 patients (24%). In 9% of cases, the Orion catheter detected far-field signals originating from the right atrium. Quantitative assessment of the created lesion revealed a significant reduction of the left atrial area having voltage >0.5 mV. A total of 29 patients (88%) remained free of symptomatic AF during a mean follow-up of 13.2 ± 3.7 months.
Conclusion: Atrial re-mapping after CB ablation by means of a HRM system improves the detection of areas of incomplete ablation, characterizes the extension of the cryo-ablated tissue and can identify abolishment of potential non-PVI related sources of AF.
Keywords: Ablation; Atrial fibrillation; Cryoballoon; High-resolution mapping; Pulmonary vein isolation.
Copyright © 2018 Elsevier B.V. All rights reserved.
Comment in
-
Making the invisible visible - Ultra-high-resolution mapping to identify residual pulmonary vein conduction immediately after cryoballoon-based pulmonary vein isolation.Int J Cardiol. 2018 Nov 1;270:180-182. doi: 10.1016/j.ijcard.2018.07.093. Epub 2018 Jul 20. Int J Cardiol. 2018. PMID: 30054145 No abstract available.