Comparison of Holladay equivalent keratometry readings and anterior corneal surface keratometry measurements in keratoconus
- PMID: 29934929
- DOI: 10.1007/s10792-018-0967-2
Comparison of Holladay equivalent keratometry readings and anterior corneal surface keratometry measurements in keratoconus
Abstract
Purpose: To compare the accuracy of the anterior corneal simulated keratometry (SimK) and the Holladay equivalent keratometry reading (EKR) provided by a Scheimpflug camera (Pentacam HR) with the keratometry (K) provided by a Placido system (T-Cone topography) in keratoconus and control eyes.
Methods: This prospective study included 40 consecutive patients with keratoconus and 40 voluntary participants with no ocular complaints. Any patients with corneal scar, corneal trauma, history of corneal surgery or contact lens usage were excluded from the study. Mean SimK and Holladay EKR measurements were taken with Pentacam HR in the 2, 3, and 4.5 mm corneal zones, and these values were compared with the T-Cone mean K value with the Placido topography system attachment on the Lenstar LS 900. Statistical analysis was performed using the paired Student's t test and the Bland-Altman analysis.
Results: A statistically significant difference was determined between the Placido K and the Scheimpflug EKR 2, 3, 4.5 mm and SimK values in the keratoconus group (p < 0.05). In the analyses which showed a difference between the SimK and Holladay EKR, it was observed that as the diameter of the corneal zone increased, the 95% LoA values were extended. No statistically significant difference was determined between the SimK and EKR 2 mm values (p > 0.05). In the control group, there was no statistically significant difference between any of the keratometric values.
Conclusion: In diseases which affect the posterior corneal surface, such as keratoconus, it is thought that because of the asymmetrical peripheral placement of the corneal apex, as the corneal diameter increases there could be an error increase of 1-3 mm in keratometric systems evaluating the anterior surface.
Keywords: Equivalent keratometry reading; Keratoconus; Placido; Scheimpflug.
Similar articles
-
An Analysis of Scheimpflug Holladay-Equivalent Keratometry Readings Following Corneal Collagen Cross-Linking.Beyoglu Eye J. 2019 Aug 8;4(2):62-68. doi: 10.14744/bej.2019.35220. eCollection 2019. Beyoglu Eye J. 2019. PMID: 35187435 Free PMC article.
-
Comparison of the accuracy of four Pentacam corneal astigmatism values in non-toric pseudophakic eyes.Graefes Arch Clin Exp Ophthalmol. 2020 Apr;258(4):795-803. doi: 10.1007/s00417-019-04585-8. Epub 2020 Jan 3. Graefes Arch Clin Exp Ophthalmol. 2020. PMID: 31900645
-
Scheimpflug camera combined with placido-disk corneal topography and optical biometry for intraocular lens power calculation.Int Ophthalmol. 2017 Aug;37(4):781-786. doi: 10.1007/s10792-016-0330-4. Epub 2016 Sep 1. Int Ophthalmol. 2017. PMID: 27586670
-
Cornea and anterior eye assessment with placido-disc keratoscopy, slit scanning evaluation topography and scheimpflug imaging tomography.Indian J Ophthalmol. 2018 Mar;66(3):360-366. doi: 10.4103/ijo.IJO_850_17. Indian J Ophthalmol. 2018. PMID: 29480244 Free PMC article. Review.
-
Scheimpflug imaging for keratoconus and ectatic disease.Indian J Ophthalmol. 2013 Aug;61(8):401-6. doi: 10.4103/0301-4738.116059. Indian J Ophthalmol. 2013. PMID: 23925323 Free PMC article. Review.
Cited by
-
Intraocular lens power calculation in patients with irregular astigmatism.Graefes Arch Clin Exp Ophthalmol. 2022 Dec;260(12):3889-3895. doi: 10.1007/s00417-022-05729-z. Epub 2022 Jul 1. Graefes Arch Clin Exp Ophthalmol. 2022. PMID: 35776172
-
Evaluation of topographic, tomographic, topometric, densitometric, and aberrometric features of cornea with pentacam HR system in subclinical keratoconus.Int Ophthalmol. 2021 May;41(5):1729-1741. doi: 10.1007/s10792-021-01732-1. Epub 2021 Mar 27. Int Ophthalmol. 2021. PMID: 33772701
-
An Analysis of Scheimpflug Holladay-Equivalent Keratometry Readings Following Corneal Collagen Cross-Linking.Beyoglu Eye J. 2019 Aug 8;4(2):62-68. doi: 10.14744/bej.2019.35220. eCollection 2019. Beyoglu Eye J. 2019. PMID: 35187435 Free PMC article.
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Miscellaneous