. 2018 Oct;289(1):255-260.
doi: 10.1148/radiol.2018180188. Epub 2018 Jun 26.

Phantom Study of In-Stent Restenosis at High-Spatial-Resolution CT

Affiliations

Phantom Study of In-Stent Restenosis at High-Spatial-Resolution CT

Hiromitsu Onishi et al. Radiology. 2018 Oct.

Abstract

Purpose To examine the diagnostic performance of high-spatial-resolution (HSR) CT with 0.25-mm section thickness for evaluating renal artery in-stent restenosis. Materials and Methods A 0.05-mm wire phantom and vessel phantoms with renal stents with in-stent stenotic sections of varying diameters were scanned with both an HSR CT scanner equipped with 160-section multi-detector rows (0.25-mm section thickness) and a conventional CT scanner. The wire phantom was used to analyze modulation transfer function (MTF). With the vessel phantoms, the error rates were calculated as the absolute difference between the measured diameters and true diameters divided by the true diameters at the narrowing sections. For qualitative evaluation, overall image quality and diagnostic accuracy for evaluating stenosis in three stages were assessed by two radiologists. Statistical analyses included the paired t test, Wilcoxon signed-rank test, and McNemar test. Results HSR CT achieved 24.3 line pairs per centimeter ± 0.5 (standard deviation) and 29.1 line pairs per centimeter ± 0.4 at 10% and 2% MTF, respectively; and conventional CT was 12.5 line pairs per centimeter ± 0.1 and 14.3 line pairs per centimeter ± 0.1 at 10% and 2% MTF, respectively. The mean error rate of the measured diameter at HSR CT (8.0% ± 5.8) was significantly lower than that at at conventional CT (16.9% ± 9.3; P < .001). Image quality at HSR CT was significantly better than that at conventional CT (P < .001), but HSR CT was not significantly superior to conventional CT in terms of diagnostic accuracy. Conclusion Compared with conventional CT, high-spatial-resolution CT achieved spatial resolutions of up to 29 line pairs per centimeter at 2% modulation transfer function and yielded improved measurement accuracy for the evaluation of in-stent restenosis in a phantom study of renal artery stents. Published under a CC BY 4.0 license.

PubMed Disclaimer

Comment in

Publication types

MeSH terms

LinkOut - more resources