Evaluation of low bone mass and prediction of fracture risk using metacarpal radiogrammetry method: a comparative study with DXA and X-ray phantom
- PMID: 29968333
- DOI: 10.1111/1756-185X.13326
Evaluation of low bone mass and prediction of fracture risk using metacarpal radiogrammetry method: a comparative study with DXA and X-ray phantom
Abstract
Objectives: (i) To predict the future risk of osteoporotic fracture in women using a simple forearm radiograph. (ii) To assess osteoporosis in southern Indian women by using radiogrammetric technique in comparison with dual-energy X-ray absorptiometry (DXA) and X-ray phantom study.
Methods: The bone mineral density (BMD) of the right proximal femur by DXA and the X-ray measurements were acquired from the right forearm. The combined cortical thickness at the second to fourth metacarpal region (M-CCT), radius (R-CCT) and ulna (U-CCT) were derived in all the studied population. The aluminium phantom study was conducted by varying the X-ray source to film distance at 100 cm and 150 cm, respectively. The feed forward back propagation neural network was used for classification of low bone mass group and normal.
Results: The combined cortical thickness of M-CCT, R-CCT and U-CCT of the total studied population was strongly correlated with DXA femur Th.BMD measurements (r = 0.77, r = 0.61 and r = 0.59 [P < 0.01]). The predicted future osteoporotic fracture risk for the low bone mass group, post-menopausal women and old-aged women population was found to be 92%, 62.8%, and 64.7%, respectively. The accuracy of neural network classifier for training set, testing set was found to be 97.5% and 87.5% in the studied population.
Conclusion: The results suggested that M-CCT and M-CCT (%) at the second metacarpal region are useful in predicting the future risk of osteoporotic fracture in women. The aluminium phantom study with an X-ray tube to film distance of 100 cm mimics an exact condition of forearm radiogrammetry.
Keywords: bone mineral densitometry; cortical thickness; dual X-ray absorptiometry; metacarpal; osteoporosis; radiogrammetry.
© 2018 Asia Pacific League of Associations for Rheumatology and John Wiley & Sons Australia, Ltd.
Similar articles
-
The role of hip and chest radiographs in osteoporotic evaluation among south Indian women population: a comparative scenario with DXA.J Endocrinol Invest. 2014 May;37(5):429-40. doi: 10.1007/s40618-014-0074-9. Epub 2014 Apr 16. J Endocrinol Invest. 2014. PMID: 24737214
-
Computed radiographic absorptiometry and morphometry in the assessment of postmenopausal bone loss.Osteoporos Int. 1996;6(1):8-13. doi: 10.1007/BF01626531. Osteoporos Int. 1996. PMID: 8845604
-
Comparison of three plain radiography methods for evaluating proximal humerus bone strength in women.J Orthop Sci. 2019 Mar;24(2):243-249. doi: 10.1016/j.jos.2018.09.020. Epub 2018 Oct 23. J Orthop Sci. 2019. PMID: 30361168
-
Bone density measurement and assessment of fracture risk.Clin Obstet Gynecol. 2013 Dec;56(4):667-76. doi: 10.1097/GRF.0b013e3182a8ad7e. Clin Obstet Gynecol. 2013. PMID: 24022502 Review.
-
Use of Trabecular Bone Score (TBS) as a Complementary Approach to Dual-energy X-ray Absorptiometry (DXA) for Fracture Risk Assessment in Clinical Practice.J Clin Densitom. 2017 Jul-Sep;20(3):334-345. doi: 10.1016/j.jocd.2017.06.019. Epub 2017 Jul 19. J Clin Densitom. 2017. PMID: 28734710 Review.
Cited by
-
Artificial intelligence, osteoporosis and fragility fractures.Curr Opin Rheumatol. 2019 Jul;31(4):368-375. doi: 10.1097/BOR.0000000000000607. Curr Opin Rheumatol. 2019. PMID: 31045948 Free PMC article. Review.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical