Nonlinear Flow Sensor Calibration with an Accurate Syringe
- PMID: 29976851
- PMCID: PMC6068951
- DOI: 10.3390/s18072163
Nonlinear Flow Sensor Calibration with an Accurate Syringe
Abstract
Flow sensors are required for monitoring patients on mechanical ventilation and in respiratory research. Proper calibration is important for ensuring accuracy and can be done with a precision syringe. This procedure, however, becomes complex for nonlinear flow sensors, which are commonly used. The objective of the present work was to develop an algorithm to allow the calibration of nonlinear flow sensors using an accurate syringe. We first noticed that a power law equation could properly fit the pressure-flow relationship of nonlinear flow sensors. We then developed a software code to estimate the parameters for this equation using a 3 L syringe (calibration syringe). Finally, we tested the performance of a calibrated flow sensor using a different 3 L syringe (testing syringe) and a commercially available spirometer. After calibration, the sensor had a bias ranging from −1.7% to 3.0% and precision from 0.012 L to 0.039 L for volumes measured with the 3 L testing syringe. Calibrated sensor performance was at least as good as the commercial sensor. This calibration procedure can be done at the bedside for both clinical and research purposes, therefore improving the accuracy of nonlinear flow sensors.
Keywords: calibration; flow sensor; mechanical ventilation; nonlinear.
Conflict of interest statement
The authors declare no conflict of interest.
Figures
References
-
- The Acute Respiratory Distress Syndrome Network The Acute Respiratory Distress Syndrome Network Ventilation with lower tidal volumes as compared with traditional tidal volumes for acute lung injury and the acute respiratory distress syndrome. N. Engl. J. Med. 2000;342:1301–1308. doi: 10.1056/NEJM200005043421801. - DOI - PubMed
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources