Use of the Airo mobile intraoperative CT system versus the O-arm for transpedicular screw fixation in the thoracic and lumbar spine: a retrospective cohort study of 263 patients
- PMID: 29979141
- DOI: 10.3171/2018.1.SPINE17927
Use of the Airo mobile intraoperative CT system versus the O-arm for transpedicular screw fixation in the thoracic and lumbar spine: a retrospective cohort study of 263 patients
Abstract
Objective: Navigation-enabling technology such as 3D-platform (O-arm) or intraoperative mobile CT (iCT-Airo) systems for use in spinal surgery has considerably improved accuracy over that of traditional fluoroscopy-guided techniques during pedicular screw positioning. In this study, the authors compared 2 intraoperative imaging systems with navigation, available in their neurosurgical unit, in terms of the accuracy they provided for transpedicular screw fixation in the thoracic and lumbar spine.
Methods: The authors performed a retrospective analysis of clinical and surgical data of 263 consecutive patients who underwent thoracic and lumbar spine screw placement in the same center. Data on 97 patients who underwent surgery with iCT-Airo navigation (iCT-Airo group) and 166 with O-arm navigation (O-arm group) were analyzed. Most patients underwent surgery for a degenerative or traumatic condition that involved thoracic and lumbar pedicle screw fixation using an open or percutaneous technique. The primary endpoint was the proportion of patients with at least 1 screw not correctly positioned according to the last intraoperative image. Secondary endpoints were the proportion of screws that were repositioned during surgery, the proportion of patients with a postoperative complication related to screw malposition, surgical time, and radiation exposure. A blinded radiologist graded screw positions in the last intraoperative image according to the Heary classification (grade 1-3 screws were considered correctly placed).
Results: A total of 1361 screws placed in 97 patients in the iCT-Airo group (503 screws) and in 166 in the O-arm group (858 screws) were graded. Of those screws, 3 (0.6%) in the iCT-Airo group and 4 (0.5%) in the O-arm group were misplaced. No statistically significant difference in final accuracy between these 2 groups or in the subpopulation of patients who underwent percutaneous surgery was found. Three patients in the iCT-Airo group (3.1%, 95% CI 0%-6.9%) and 3 in the O-arm group (1.8%, 95% CI 0%-4.0%) had a misplaced screw (Heary grade 4 or 5). Seven (1.4%) screws in the iCT-Airo group and 37 (4.3%) in the O-arm group were repositioned intraoperatively (p = 0.003). One patient in the iCT-Airo group and 2 in the O-arm group experienced postoperative neurological deficits related to hardware malposition. The mean surgical times in both groups were similar (276 [iCT-Airo] and 279 [O-arm] minutes). The mean exposure to radiation in the iCT-Airo group was significantly lower than that in the O-arm group (15.82 vs 19.12 mSv, respectively; p = 0.02).
Conclusions: Introduction of a mobile CT scanner reduced the rate of screw repositioning, which enhanced patient safety and diminished radiation exposure for patients, but it did not improve overall accuracy compared to that of a mobile 3D platform.
Keywords: 3D imaging; IQR = interquartile range; computed tomography; iCT = intraoperative CT; intraoperative; lumbar; navigation; pedicle screw; spine; surgical technique; thoracic.
Similar articles
-
Use of Intraoperative CT Improves Accuracy of Spinal Navigation During Screw Fixation in Cervico-thoracic Region.Spine (Phila Pa 1976). 2021 Apr 15;46(8):530-537. doi: 10.1097/BRS.0000000000003827. Spine (Phila Pa 1976). 2021. PMID: 33273435
-
Prospective Comparative Study in Spine Surgery Between O-Arm and Airo Systems: Efficacy and Radiation Exposure.World Neurosurg. 2018 Oct;118:e175-e184. doi: 10.1016/j.wneu.2018.06.148. Epub 2018 Jul 3. World Neurosurg. 2018. PMID: 30257292
-
Intraoperative Computed Tomography Versus 3D C-Arm Imaging for Navigated Spinal Instrumentation.Spine (Phila Pa 1976). 2018 Mar 1;43(5):370-377. doi: 10.1097/BRS.0000000000002173. Spine (Phila Pa 1976). 2018. PMID: 28368989
-
How Do Robotics and Navigation Facilitate Minimally Invasive Spine Surgery? A Case Series and Narrative Review.Neurosurgery. 2025 Mar 1;96(3S):S84-S93. doi: 10.1227/neu.0000000000003352. Epub 2025 Feb 14. Neurosurgery. 2025. PMID: 39950788 Review.
-
Robot-assisted percutaneous pedicle screw placement accuracy compared with alternative guidance in lateral single-position surgery: a systematic review and meta-analysis.J Neurosurg Spine. 2023 Jun 23;39(4):443-451. doi: 10.3171/2023.3.SPINE2329. Print 2023 Oct 1. J Neurosurg Spine. 2023. PMID: 37382304
Cited by
-
Role of intraoperative computed tomography scanner in modern neurosurgery - An early experience.Surg Neurol Int. 2020 Aug 15;11:247. doi: 10.25259/SNI_303_2020. eCollection 2020. Surg Neurol Int. 2020. PMID: 32905376 Free PMC article.
-
Technical assessment of a mobile CT scanner for image-guided brachytherapy.J Appl Clin Med Phys. 2019 Oct;20(10):187-200. doi: 10.1002/acm2.12738. Epub 2019 Oct 2. J Appl Clin Med Phys. 2019. PMID: 31578811 Free PMC article.
-
Feasibility and accuracy evaluation of novel 2D instant navigation system on spinal surgery - a preclinical study.BMC Musculoskelet Disord. 2025 May 23;26(1):510. doi: 10.1186/s12891-025-08769-8. BMC Musculoskelet Disord. 2025. PMID: 40410739 Free PMC article.
-
Comparison of pedicle screw placement accuracy between two types of imaging support (Artis Zeego versus two-dimensional fluoroscopy): a cross-sectional observational study.BMC Musculoskelet Disord. 2022 Jul 5;23(1):644. doi: 10.1186/s12891-022-05602-4. BMC Musculoskelet Disord. 2022. PMID: 35790951 Free PMC article.
-
Robotics and navigation in spine surgery: A narrative review.J Orthop. 2023 Aug 22;44:36-46. doi: 10.1016/j.jor.2023.08.007. eCollection 2023 Oct. J Orthop. 2023. PMID: 37664556 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical