. 2019 Mar;23(3):1263-1270.
doi: 10.1007/s00784-018-2558-z. Epub 2018 Jul 8.

Diagnosis of vertical root fracture in teeth close and distant to implant: an in vitro study to assess the influence of artifacts produced in cone beam computed tomography

Affiliations

Diagnosis of vertical root fracture in teeth close and distant to implant: an in vitro study to assess the influence of artifacts produced in cone beam computed tomography

Deborah Queiroz Freitas et al. Clin Oral Investig. 2019 Mar.

Abstract

Objectives: To evaluate the influence of artifacts produced by zirconium implant on the diagnosis of vertical root fracture (VRF) in teeth close and distant to the implant in cone beam computed tomography (CBCT) images. We also determined if kilovoltage (kVp) and metal artifact reduction (MAR) tool could influence this diagnosis.

Materials and methods: Twenty single-root teeth were divided in control and fractured groups (n = 10). The teeth were randomly positioned in the first and second and right and left pre-molar alveoli of a dry human mandible. CBCT exams were acquired using a ProMax 3D unit with varying kVp (70, 80, or 90 kVp), with or without MAR, and with and without a zirconium implant placed in the alveolus of first right molar. The images were evaluated by five observers. The area under the receiver operating characteristic curve (ROC), sensitivity, and specificity were calculated and compared by analysis of variance with a significance level of 5%.

Results: In general, ROC and sensitivity were not affected by the factors studied (p > 0.05). The main effects occurred in specificity; when implant was used without MAR, the values were lower for tooth 45 for all kVps (p = 0.0001).

Conclusions: Artifacts produced in the vicinity of teeth with suspected VRF impair the diagnosis by decreasing the specificity, because they can mimic the VRF line generating false positives. However, MAR improves the specificity, being its use recommended when metallic objects are present near teeth with suspected VRF.

Clinical relevance: Since nowadays, many patients who undergo CBCT show implants and they definitively produce artifacts, it is important to evaluate the influence of such artifacts in the diagnosis of teeth that are close to the generator-artifact object.

Keywords: Artifact; Artifact reduction tool; Cone beam computed tomography; Kilovoltage; Vertical root fracture.

PubMed Disclaimer

References

    1. Clin Oral Implants Res. 2010 Jan;21(1):100-7 - PubMed
    1. J Endod. 2011 Jun;37(6):768-72 - PubMed
    1. Clin Oral Implants Res. 2013 Aug;24 Suppl A100:94-9 - PubMed
    1. J Endod. 2012 Oct;38(10):1344-7 - PubMed
    1. Dentomaxillofac Radiol. 2013;42(5):20120245 - PubMed

Substances

LinkOut - more resources