Design and Functional Evaluation of a Dexterous Myoelectric Hand Prosthesis With Biomimetic Tactile Sensor
- PMID: 29985148
- DOI: 10.1109/TNSRE.2018.2844807
Design and Functional Evaluation of a Dexterous Myoelectric Hand Prosthesis With Biomimetic Tactile Sensor
Abstract
This paper presents the design, tactile sensor, characterization, and control system of a new dexterous myoelectric hand prosthesis to overcome the limitations of state-of-the-art myoelectric prostheses (e.g., limited functionality, controllability, and sensory feedback). Our dexterous myoelectric hand allows independent finger movement and thumb abduction/adduction, with a motor for each finger and an additional motor for the thumb (i.e., six total motors). Each fingertip has a biomimetic tactile sensor with 13 tactile units, each of which can detect normal and tangential forces. The hand controller uses an electromyography pattern recognition controller and a tactile sensor feedback-based grasping controller to automatically and dynamically adjust the finger grasp force to prevent objects from slipping. This closed-loop controller structure will allow users to safely and effectively grasp complex objects with varying densities and shapes. In addition, the electronic hardware is integrated into the hand, and the pattern recognition controller can be implemented in the hand embedded system.
Similar articles
-
The S-Finger: A Synergetic Externally Powered Digit With Tactile Sensing and Feedback.IEEE Trans Neural Syst Rehabil Eng. 2018 Jun;26(6):1264-1271. doi: 10.1109/TNSRE.2018.2829183. IEEE Trans Neural Syst Rehabil Eng. 2018. PMID: 29877851
-
Closed-Loop Multi-Amplitude Control for Robust and Dexterous Performance of Myoelectric Prosthesis.IEEE Trans Neural Syst Rehabil Eng. 2020 Feb;28(2):498-507. doi: 10.1109/TNSRE.2019.2959714. Epub 2019 Dec 13. IEEE Trans Neural Syst Rehabil Eng. 2020. PMID: 31841418
-
Cognitive vision system for control of dexterous prosthetic hands: experimental evaluation.J Neuroeng Rehabil. 2010 Aug 23;7:42. doi: 10.1186/1743-0003-7-42. J Neuroeng Rehabil. 2010. PMID: 20731834 Free PMC article.
-
Improving the functionality, robustness, and adaptability of myoelectric control for dexterous motion restoration.Exp Brain Res. 2019 Feb;237(2):291-311. doi: 10.1007/s00221-018-5441-x. Epub 2018 Nov 30. Exp Brain Res. 2019. PMID: 30506366 Review.
-
Myoelectric training systems.Expert Rev Med Devices. 2011 Sep;8(5):581-9. doi: 10.1586/erd.11.23. Expert Rev Med Devices. 2011. PMID: 22026623 Review.
Cited by
-
Recent Developments in Prosthesis Sensors, Texture Recognition, and Sensory Stimulation for Upper Limb Prostheses.Ann Biomed Eng. 2021 Jan;49(1):57-74. doi: 10.1007/s10439-020-02678-8. Epub 2020 Nov 2. Ann Biomed Eng. 2021. PMID: 33140242 Review.
-
Application of multimodal perception scenario construction based on IoT technology in university music teaching.PeerJ Comput Sci. 2023 Oct 25;9:e1602. doi: 10.7717/peerj-cs.1602. eCollection 2023. PeerJ Comput Sci. 2023. PMID: 38077533 Free PMC article.
-
A Piezoresistive Tactile Sensor for a Large Area Employing Neural Network.Sensors (Basel). 2018 Dec 21;19(1):27. doi: 10.3390/s19010027. Sensors (Basel). 2018. PMID: 30577675 Free PMC article.
-
The Latest Research Progress on Bionic Artificial Hands: A Systematic Review.Micromachines (Basel). 2024 Jul 8;15(7):891. doi: 10.3390/mi15070891. Micromachines (Basel). 2024. PMID: 39064402 Free PMC article. Review.
-
Designing Prosthetic Hands With Embodied Intelligence: The KIT Prosthetic Hands.Front Neurorobot. 2022 Mar 10;16:815716. doi: 10.3389/fnbot.2022.815716. eCollection 2022. Front Neurorobot. 2022. PMID: 35355833 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources