Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2019 May;43(5):1097-1105.
doi: 10.1007/s00264-018-4054-0. Epub 2018 Jul 19.

Intraprosthetic dislocation of dual mobility total hip arthroplasty: still occurring?

Affiliations
Comparative Study

Intraprosthetic dislocation of dual mobility total hip arthroplasty: still occurring?

Thomas Neri et al. Int Orthop. 2019 May.

Erratum in

Abstract

Purpose: The objective was to identify predictive factors for intraprosthetic dislocation (IPD) and to understand how improvements in dual mobility cups (DMC) have helped to reduce dramatically the occurrence of this complication.

Methods: DM mobile inserts retrieved from 93 hips were divided into three groups: first-generation DMC with IPD ("firstDMC-IPD"), first-generation DMC with over 15 years of implantation without IPD ("firstDMC-noIPD"), and latest-generation DMC ("newDMC"). The predictive factors for IPD based on clinical, prosthetic, radiological and intraoperative characteristics were analysed by multivariate analysis. The surface of each retrieved mobile insert was analysed using three-dimensional CT scan in order to compare their rim wear.

Results: Three predictive factors for IPD were found: a high BMI, a wide rough stem neck and a large cup size. Wear of the firstDMC-noIPD inserts was significantly less than those of firstDMC-IPD inserts and significantly more than those of newDMC inserts. For the firstDMC-IPD inserts, the rim's outer surface wear was significantly greater than the rim's inner surface wear.

Conclusions: IPD is a specific complication related to wear of the DM mobile insert due to failure of the liner's retaining rim, especially from the rim's outer surface. This long-term issue is different to the early traumatic complication, which can happen after an attempt at closed reduction of a DM THA dislocation. Recent modifications in the design and the coating of contemporary DMC and femoral stems, as well as improvements in the mobile insert itself, seem to corroborate our assumptions about the IPD mechanism and contribute to the quasi-disappearance of this complication.

Keywords: Dual mobility; Intraprosthetic dislocation; Total hip arthroplasty.

PubMed Disclaimer

References

    1. J Arthroplasty. 2012 Mar;27(3):493.e1-3 - PubMed
    1. Int Orthop. 2018 Jan;42(1):41-47 - PubMed
    1. Int Orthop. 2017 Mar;41(3):611-618 - PubMed
    1. J Arthroplasty. 2016 Aug;31(8):1828-35 - PubMed
    1. Clin Orthop Relat Res. 2012 Oct;470(10):2810-8 - PubMed

Publication types