. 2018 Sep;160(9):1755-1760.
doi: 10.1007/s00701-018-3642-5. Epub 2018 Jul 28.

Loop microcatheter technique for coil embolization of paraclinoid aneurysms

Affiliations

Loop microcatheter technique for coil embolization of paraclinoid aneurysms

Zhen Yu Jia et al. Acta Neurochir (Wien). 2018 Sep.

Abstract

Background: Although paraclinoid aneurysms do not exhibit a high risk of rupture, coil embolization is not always easy because of unstable microcatheter position. We present a technique that allows a stable microcatheter position for coil embolization of paraclinoid aneurysms.

Methods: We enrolled 34 consecutive patients who underwent coil embolization for paraclinoid aneurysms. A loop of distal microcatheter was shaped based on three-dimensional rotational angiography. The basic concept is to keep the proximal loop abutting the opposite wall of the aneurysm while using the distal loop for coiling. Then, a proximal curve was made to accommodate the shape of the carotid siphon, which may decide the direction of the loop. Stent-assisted coil embolization was performed in 19 wide-necked aneurysms. Immediate radiological outcomes were analyzed with Raymond classification and clinical outcomes were evaluated with modified Rankin Scale (mRS) scores.

Results: Satisfactory occlusion of aneurysm was achieved in 94.1% (32/34) of patients with a Raymond score of 1 or 2. Packing density of ≥ 31% was achieved in 71% (24/34) of patients. No significant differences were observed between stent-assisted coiling and coiling-only groups. Follow-up magnetic resonance angiography and/or angiogram showed stable coil position, except in one patient with tiny recurrence (from Raymond scores 1 to 2) that did not require retreatment at the 6-month follow-up. mRS scores of 0-1 were obtained in all patients at 6 months.

Conclusions: Loop microcatheter technique allowed safe and stable coil packing for paraclinoid aneurysms. The same procedural concept is also being used for aneurysms in other vascular territories.

Keywords: Aneurysm; Coiling; Embolization; Microcatheter shaping.

PubMed Disclaimer

Publication types