Gaitography on lower-limb amputees: Repeatability and between-methods agreement
- PMID: 30101682
- DOI: 10.1177/0309364618791618
Gaitography on lower-limb amputees: Repeatability and between-methods agreement
Abstract
Background:: Gaitography is gait parametrization from center-of-pressure trajectories of walking on an instrumented treadmill. Gaitograms may be useful for prosthetic gait analyses, as they can be rapidly and unobtrusively collected over multiple gait cycles without constraining foot placement. However, its reliability must still be established for prosthetic gait.
Objectives:: To evaluate (a) within-method test-retest repeatability and (b) between-methods agreement for temporal gait events (foot contact, foot off) and gait characteristics (e.g. step times, single-support duration).
Study design:: Cohort study with repeated measurements.
Methods:: Ten male proficient prosthetic walkers with a unilateral trans-femoral or trans-tibial amputation were equipped with a pressure-insole system and were invited to walk on separate days on an instrumented treadmill.
Results:: We found better between-methods reproducibility than within-method repeatability in temporal gait characteristics. Step times, stride times, and foot-contact events matched well between the two methods. In contrast, insole-based foot-off events were detected one-to-two samples earlier. Likewise, a similar bias was observed for temporal gait characteristics that incorporated foot-off events.
Conclusion:: Notwithstanding small systematic biases, the good between-methods agreement indicates that temporal gait characteristics may be determined interchangeably with gaitograms and insoles in persons with a prosthesis. However, the relatively poorer test-retest repeatability hinders longitudinal assessments with either method.
Clinical relevance:: Clinical practice could potentially benefit from gaitography as an efficient, unobtrusive, easy to use, automatized, and patient-friendly means to objectively parametrize prosthetic gait, with immediate availability of test results allowing for prompt clinical decision-making. Temporal gait parameters demonstrate good between-methods agreement, but poorer within-method repeatability hinders detecting prosthetic gait changes.
Keywords: Gait; amputees; center of pressure; online event detection; treadmill test.
Similar articles
-
Gaitography applied to prosthetic walking.Med Biol Eng Comput. 2014 Nov;52(11):963-969. doi: 10.1007/s11517-014-1195-1. Epub 2014 Sep 24. Med Biol Eng Comput. 2014. PMID: 25249276
-
Long-distance walking effects on trans-tibial amputees compensatory gait patterns and implications on prosthetic designs and training.Gait Posture. 2012 Feb;35(2):328-33. doi: 10.1016/j.gaitpost.2011.10.004. Epub 2011 Nov 4. Gait Posture. 2012. PMID: 22055554
-
Evaluating asymmetry in prosthetic gait with step-length asymmetry alone is flawed.Gait Posture. 2012 Mar;35(3):446-51. doi: 10.1016/j.gaitpost.2011.11.005. Epub 2011 Dec 6. Gait Posture. 2012. PMID: 22153771
-
The effect of anti-gravity treadmill training for prosthetic rehabilitation of a case with below-knee amputation.Prosthet Orthot Int. 2015 Dec;39(6):502-6. doi: 10.1177/0309364614532866. Epub 2014 May 12. Prosthet Orthot Int. 2015. PMID: 24821715 Review.
-
Prosthetic feet: state-of-the-art review and the importance of mimicking human ankle-foot biomechanics.Disabil Rehabil Assist Technol. 2009 Mar;4(2):65-75. doi: 10.1080/17483100802715092. Disabil Rehabil Assist Technol. 2009. PMID: 19253096 Review.
Cited by
-
Estimation of gait parameters using leg velocity for amputee population.PLoS One. 2022 May 13;17(5):e0266726. doi: 10.1371/journal.pone.0266726. eCollection 2022. PLoS One. 2022. PMID: 35560138 Free PMC article.
-
Distinguishing retrospective fallers from nonfallers in people who use a unilateral lower-limb prosthesis.Prosthet Orthot Int. 2025 Mar 24:10.1097/PXR.0000000000000439. doi: 10.1097/PXR.0000000000000439. Online ahead of print. Prosthet Orthot Int. 2025. PMID: 40126880
-
Consistent inconsistencies in braking: a spatial analysis.Interface Focus. 2021 Aug 13;11(5):20200058. doi: 10.1098/rsfs.2020.0058. eCollection 2021 Oct 6. Interface Focus. 2021. PMID: 34938429 Free PMC article.
-
Centre of Pressure Estimation during Walking Using Only Inertial-Measurement Units and End-To-End Statistical Modelling.Sensors (Basel). 2020 Oct 28;20(21):6136. doi: 10.3390/s20216136. Sensors (Basel). 2020. PMID: 33126671 Free PMC article.
-
Reliability and Validity of Running Cadence and Stance Time Derived from Instrumented Wireless Earbuds.Sensors (Basel). 2021 Nov 30;21(23):7995. doi: 10.3390/s21237995. Sensors (Basel). 2021. PMID: 34883999 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical