Accuracy of cone-beam computed tomography, dental magnetic resonance imaging, and intraoral radiography for detecting peri-implant bone defects at single zirconia implants-An in vitro study
- PMID: 30112833
- DOI: 10.1111/clr.13348
Accuracy of cone-beam computed tomography, dental magnetic resonance imaging, and intraoral radiography for detecting peri-implant bone defects at single zirconia implants-An in vitro study
Abstract
Objectives: To evaluate the diagnostic value of cone-beam computed tomography (CBCT), intraoral radiography (IR), and dental magnetic resonance imaging (dMRI) for detecting and classifying peri-implant bone defects at zirconia implants.
Materials and methods: Forty-eight zirconia implants were inserted in bovine ribs, 24 of which had standardized defects (1-wall, 2-wall, 3-wall, 4-wall) in two sizes (1 and 3 mm). CBCT, IR, and dMRI were performed and analyzed twice by four readers unaware of the nature of the defects. Cohen's and Fleiss' kappa (κ), sensitivity, and specificity were calculated for the presence/absence of bone defects, defect size, and defect type. Cochran's Q-test with post hoc McNemar was used to test for statistical differences.
Results: A high intra- and inter-reader reliability (κ range: 0.832-1) and sensitivity/specificity (IR: 0.97/0.96; CBCT: 0.99/1; dMRI: 1/0.99) for bone defect detection were observed for all three imaging methods. For defect type classification, intra- (κ range: 0.505-0.778) and inter-reader (κ: 0.411) reliability of IR were lower compared to CBCT (κ range intrareader: 0.667-0.889; κ inter-reader: 0.629) and dMRI (κ range intrareader: 0.61-0.832; κ inter-reader: 0.712). The sensitivity for correct defect type classification was not significantly different for CBCT (0.81) and dMRI (0.83; p = 1), but was significantly lower for IR (0.68; vs. CBCT p = 0.003; vs. dMRI p = 0.004). The sensitivity advantage of CBCT and dMRI for defect classification was smaller for 1-mm defects (CBCT/dMRI/IR: 0.68/0.72/0.63, no significant difference) than for 3-mm defects (CBCT/dMRI/IR: 0.95/0.94/0.74; CBCT vs. IR p = 0.0001; dMRI vs. IR p = 0.003).
Conclusion: Within the limitations of an in vitro study, IR can be recommended as the initial imaging method for evaluating peri-implant bone defects at zirconia implants. CBCT provides higher diagnostic accuracy of defect classification at the expense of higher cost and radiation dose. Dental MRI may be a promising imaging method for evaluating peri-implant bone defects at zirconia implants in the future.
Keywords: CBCT; MRI; dental implants; peri-implantitis; radiography; tooth.
© 2018 John Wiley & Sons A/S. Published by John Wiley & Sons Ltd.
Similar articles
-
Reliability and accuracy of intraoral radiography, cone beam CT, and dental MRI for evaluation of peri-implant bone lesions at zirconia implants - an ex vivo feasibility study.J Dent. 2023 Mar;130:104422. doi: 10.1016/j.jdent.2023.104422. Epub 2023 Jan 14. J Dent. 2023. PMID: 36649822
-
Assessment of peri-implant defects at titanium and zirconium dioxide implants by means of periapical radiographs and cone beam computed tomography: An in-vitro examination.Clin Oral Implants Res. 2018 Dec;29(12):1195-1201. doi: 10.1111/clr.13383. Epub 2018 Nov 22. Clin Oral Implants Res. 2018. PMID: 30387207
-
Detection of peri-implant bone defects with different radiographic techniques - a human cadaver study.Clin Oral Implants Res. 2016 May;27(5):529-34. doi: 10.1111/clr.12619. Epub 2015 Jun 7. Clin Oral Implants Res. 2016. PMID: 26059443
-
Comparative analysis of imaging techniques for diagnostic accuracy of peri-implant bone defects: a meta-analysis.Oral Surg Oral Med Oral Pathol Oral Radiol. 2017 Oct;124(4):432-440.e5. doi: 10.1016/j.oooo.2017.06.119. Epub 2017 Jun 23. Oral Surg Oral Med Oral Pathol Oral Radiol. 2017. PMID: 28743664 Review.
-
Diagnostic performance of cone beam computed tomography in assessing peri-implant bone loss: A systematic review.Clin Oral Implants Res. 2018 May;29(5):443-464. doi: 10.1111/clr.13143. Epub 2018 Mar 26. Clin Oral Implants Res. 2018. PMID: 29578266
Cited by
-
Efficacy of two radiographic algorithms for detection of peri-implant bone defects on cone-beam computed tomography scans.BMC Oral Health. 2025 Jan 7;25(1):39. doi: 10.1186/s12903-024-05397-x. BMC Oral Health. 2025. PMID: 39773233 Free PMC article.
-
Retrospective Analysis of Artifacts in Cone Beam Computed Tomography Images Used to Diagnose Chronic Rhinosinusitis.Diagnostics (Basel). 2023 Apr 22;13(9):1511. doi: 10.3390/diagnostics13091511. Diagnostics (Basel). 2023. PMID: 37174903 Free PMC article.
-
CBCT Validation Study on Intraclass Correlation for Linear Measurements in Peri-implantitis: an Observational Study.J Oral Maxillofac Res. 2025 Mar 31;16(1):e4. doi: 10.5037/jomr.2025.16104. eCollection 2025 Jan-Mar. J Oral Maxillofac Res. 2025. PMID: 40443422 Free PMC article.
-
Detection of peri-implant bone defects using cone-beam computed tomography and digital periapical radiography with parallel and oblique projection.Imaging Sci Dent. 2019 Dec;49(4):265-272. doi: 10.5624/isd.2019.49.4.265. Epub 2019 Dec 24. Imaging Sci Dent. 2019. PMID: 31915611 Free PMC article.
-
In Vitro Handling Characteristics of a Particulate Bone Substitute for Ridge Preservation Procedures.Materials (Basel). 2024 Jan 8;17(2):313. doi: 10.3390/ma17020313. Materials (Basel). 2024. PMID: 38255481 Free PMC article.
MeSH terms
Substances
Grants and funding
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical