Usability Testing of a Reusable Pulse Oximeter Probe Developed for Health-Care Workers Caring for Children < 5 Years Old in Low-Resource Settings
- PMID: 30141389
- PMCID: PMC6159595
- DOI: 10.4269/ajtmh.18-0016
Usability Testing of a Reusable Pulse Oximeter Probe Developed for Health-Care Workers Caring for Children < 5 Years Old in Low-Resource Settings
Abstract
Hypoxemia measured by pulse oximetry predicts child pneumonia mortality in low-resource settings (LRS). Existing pediatric oximeter probes are prohibitively expensive and/or difficult to use, limiting LRS implementation. Using a human-centered design, we developed a low-cost, reusable pediatric oximeter probe for LRS health-care workers (HCWs). Here, we report probe usability testing. Fifty-one HCWs from Malawi, Bangladesh, and the United Kingdom participated, and seven experts provided reference measurements. Health-care workers and experts measured the peripheral arterial oxyhemoglobin saturation (SpO2) independently in < 5 year olds. Health-care worker measurements were classed as successful if recorded in 5 minutes (or shorter) and physiologically appropriate for the child, using expert measurements as the reference. All expert measurements were considered successful if obtained in < 5 minutes. We analyzed the proportion of successful SpO2 measurements obtained in < 1, < 2, and < 5 minutes and used multivariable logistic regression to predict < 1 minute successful measurements. We conducted four testing rounds with probe modifications between rounds, and obtained 1,307 SpO2 readings. Overall, 67% (876) of measurements were successful and achieved in < 1 minute, 81% (1,059) < 2 minutes, and 90% (1,181) < 5 minutes. Compared with neonates, increasing age (infant adjusted odds ratio [aOR]; 1.87, 95% confidence interval [CI]: 1.16, 3.02; toddler aOR: 4.33, 95% CI: 2.36, 7.97; child aOR; 3.90, 95% CI: 1.73, 8.81) and being asleep versus being calm (aOR; 3.53, 95% CI: 1.89, 6.58), were associated with < 1 minute successful measurements. In conclusion, we designed a novel, reusable pediatric oximetry probe that was effectively used by LRS HCWs on children. This probe may be suitable for LRS implementation.
Figures



Similar articles
-
Opportunities and barriers in paediatric pulse oximetry for pneumonia in low-resource clinical settings: a qualitative evaluation from Malawi and Bangladesh.BMJ Open. 2018 Jan 30;8(1):e019177. doi: 10.1136/bmjopen-2017-019177. BMJ Open. 2018. PMID: 29382679 Free PMC article.
-
Performance of a novel reusable pediatric pulse oximeter probe.Pediatr Pulmonol. 2019 Jul;54(7):1052-1059. doi: 10.1002/ppul.24295. Epub 2019 Mar 25. Pediatr Pulmonol. 2019. PMID: 30912314 Free PMC article. Clinical Trial.
-
Effect of probe design on accuracy and reliability of pulse oximetry in pediatric patients.J Clin Anesth. 1999 Jun;11(4):323-7. doi: 10.1016/s0952-8180(99)00053-7. J Clin Anesth. 1999. PMID: 10470635
-
Accuracy of laminated disposable pulse-oximeter sensors.Respir Care. 1995 Jul;40(7):728-33. Respir Care. 1995. PMID: 10144067 Review.
-
Advanced Uses of Pulse Oximetry for Monitoring Mechanically Ventilated Patients.Anesth Analg. 2017 Jan;124(1):62-71. doi: 10.1213/ANE.0000000000001283. Anesth Analg. 2017. PMID: 27183375 Review.
Cited by
-
Designing a Smartphone-Based Pulse Oximeter for Children in South Africa (Phefumla Project): Qualitative Analysis of Human-Centered Design Workshops With Health Care Workers.JMIR Hum Factors. 2024 May 30;11:e54983. doi: 10.2196/54983. JMIR Hum Factors. 2024. PMID: 38825834 Free PMC article.
-
Usability of pulse oximeters used by community health and primary care workers as screening tools for severe illness in children under five in low resource settings: A cross-sectional study in Cambodia, Ethiopia, South Sudan, and Uganda.PLOS Glob Public Health. 2023 Jul 18;3(7):e0001800. doi: 10.1371/journal.pgph.0001800. eCollection 2023. PLOS Glob Public Health. 2023. PMID: 37463164 Free PMC article.
-
Introducing pulse oximetry for outpatient management of childhood pneumonia: An implementation research adopting a district implementation model in selected rural facilities in Bangladesh.EClinicalMedicine. 2022 Jun 29;50:101511. doi: 10.1016/j.eclinm.2022.101511. eCollection 2022 Aug. EClinicalMedicine. 2022. PMID: 35795715 Free PMC article.
-
Evaluation of the routine implementation of pulse oximeters into integrated management of childhood illness (IMCI) guidelines at primary health care level in West Africa: the AIRE mixed-methods research protocol.BMC Health Serv Res. 2022 Dec 24;22(1):1579. doi: 10.1186/s12913-022-08982-4. BMC Health Serv Res. 2022. PMID: 36566173 Free PMC article.
-
Pulse oximetry in pediatric care: Balancing advantages and limitations.World J Clin Pediatr. 2024 Sep 9;13(3):96950. doi: 10.5409/wjcp.v13.i3.96950. eCollection 2024 Sep 9. World J Clin Pediatr. 2024. PMID: 39350904 Free PMC article.
References
-
- WHO , 2013. Guidelines for the Management of Common Childhood Illnesses: Second Edition. Geneva, Switzerland: World Health Organization.
-
- Duke T, Wandi F, Jonathan M, Matai S, Kaupa M, Saavu M, Subhi R, Peel D, 2008. Improved oxygen systems for childhood pneumonia: a multihospital effectiveness study in Papua New Guinea. Lancet 372: 1328–1333. - PubMed
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical