DNA breaks induced by iodine-containing contrast medium in radiodiagnostics: a problem of tungsten?

Affiliations
Editorial

DNA breaks induced by iodine-containing contrast medium in radiodiagnostics: a problem of tungsten?

Mélanie L Ferlazzo et al. Eur Radiol Exp. .

Abstract

Iodine-containing contrast media (ICM) are extensively used to improve image quality and information content in x-ray-based examinations, particularly in computed tomography (CT). In parallel, there is increasing evidence that the use of ICM during CT sessions is associated with deoxyribonucleic acid (DNA) breaks that may influence the estimation of the risks linked to x-ray exposure. Why has iodine been preferred to any other heavy elements to enhance contrast in radiodiagnostics? How to understand such DNA breaks effect? We searched for the answers in the early times of x-ray medical use. It appeared that the maximal ratio between the relative iodine and water mass energy absorption coefficients is reached in the range of 40-60 keV, which defines the energy range in which the dose is preferentially absorbed by ICM. This range does not correspond to the K-edge of iodine but to that of tungsten, the major component of the x-ray tube anode of CT scanners. At such energy, radiolysis of the ICM produces sodium or potassium iodide that prevents a normal DNA breaks repair and influences the individual response to x-ray low-dose. Both contrast enhancement and DNA breaks effect may therefore be caused by tungsten of the anodes of x-ray tubes.

Keywords: Contrast media; DNA; Iodine; Radiation tolerance; Tomography, x-ray computed.

PubMed Disclaimer

Conflict of interest statement

Not applicableNot applicableThe PhD thesis of CD was supported by Fibermetrix (Strasbourg, France) and Neolys Diagnostics (Lyon, France). The remaining authors declare no competing interests.Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.

Figures

Fig. 1
Fig. 1
Schematic representation of the x-ray emission and absorption spectra related to the use of ICM during a CT session. a Relative emission x-ray spectrum of a typical 120 kVp CT scan. The arrow indicates the 33.2 keV component of the spectrum that would correspond to the K-edge of iodine. Iodine b and water c mass energy coefficients μen/ρ expressed in cm2.g− 1 as a function of energy. d Relative energy absorption of iodine in water (normalised ratio between the relative iodine and water mass energy absorption coefficients μen/ρ) as a function of energy. e Product of the relative emission intensity of a typical 120 kVp CT x-ray tube and the relative absorption of x-ray by ICM

References

    1. Bae KT. Intravenous contrast medium administration and scan timing at CT: considerations and approaches. Radiology. 2010;256:32–61. doi: 10.1148/radiol.10090908. - DOI - PubMed
    1. Nyman U, Aspelin P, Almen T. Iodinated contrast media: a semantic somersault. Radiology. 2015;275:934. doi: 10.1148/radiol.2015150066. - DOI - PubMed
    1. Haschek E, Lindenthal O. Contribution to the practical use of photography according to Röntgen. Wien Klin Wochenschr. 1896;9:63.
    1. Destot E, Bérard L. Société des sciences médicales de Lyon - Séance du 11 novembre 1896. Lyon Médical. 1896;88:392.
    1. Mould RF. A century of X-rays and radioactivity in medicine. London: Institute of Physics Publishing, CRC Press; 1993.