Noninvasive Transorbital Assessment of the Optic Nerve Sheath in Children: Relationship Between Optic Nerve Sheath Diameter, Deformability Index, and Intracranial Pressure
- PMID: 30169680
- DOI: 10.1093/ons/opy231
Noninvasive Transorbital Assessment of the Optic Nerve Sheath in Children: Relationship Between Optic Nerve Sheath Diameter, Deformability Index, and Intracranial Pressure
Abstract
Background: Measurement of optic nerve sheath diameter (ONSD) is a promising technique for noninvasive assessment of intracranial pressure (ICP), but has certain limitations. A recent study showed that the deformability index (DI), a dynamic parameter quantifying the pulsatile nature of the optic nerve sheath, could differentiate between patients with high vs normal ICP.
Objective: To further evaluate the diagnostic accuracy of the DI, when interpreted together with ONSD.
Methods: This prospective study included children undergoing invasive ICP measurement as part of their clinical management. Ultrasound images of the optic nerve sheath were acquired prior to measuring ICP, the images were further processed to obtain the DI. Patients were dichotomized into high (≥20 mm Hg) or normal ICP groups and compared using the Mann-Whitney U-test. Diagnostic accuracy was described using area under the receiver operating characteristic curve (AUC), sensitivity and specificity, correlation between DI, ONSD, and ICP was investigated using linear regression.
Results: A total of 28 patients were included (19 high ICP). The DI was lower in the high ICP group (0.105 vs 0.28, P = .001). AUC was 0.87, and a cut-off value of DI ≤ 0.185 demonstrated sensitivity of 89.5% and specificity of 88.9%. Diagnostic accuracy improved when combining DI with ONSD (AUC 0.98, sensitivity 94.7%, specificity 88.9%) and correlation with ICP improved when combined analysis of DI and ONSD was performed (Pearson correlation coefficient: 0.82 vs 0.42, respectively, P = .012).
Conclusion: The DI was significantly lower for patients with high vs normal ICP. This relationship improved further when the DI and ONSD were interpreted together.
Keywords: Deformability index; Noninvasive ICP; Optic nerve sheath; Ultrasound.
Copyright © 2018 by the Congress of Neurological Surgeons.
Comment in
-
Letter: Noninvasive Transorbital Assessment of the Optic Nerve Sheath in Children: Relationship Between Optic Nerve Sheath Diameter, Deformability Index, and Intracranial Pressure.Oper Neurosurg. 2020 Jan 1;18(1):E27-E28. doi: 10.1093/ons/opz284. Oper Neurosurg. 2020. PMID: 31538185 No abstract available.
-
In Reply: Noninvasive Transorbital Assessment of the Optic Nerve Sheath in Children: Relationship Between Optic Nerve Sheath Diameter, Deformability Index, and Intracranial Pressure.Oper Neurosurg. 2020 Jan 1;18(1):E29. doi: 10.1093/ons/opz286. Oper Neurosurg. 2020. PMID: 31584073 No abstract available.
Similar articles
-
Diagnostic accuracy of ultrasonographic optic nerve sheath diameter compared to invasive ICP monitoring for intracranial hypertension in pediatric patients.Pediatr Radiol. 2025 Jul;55(8):1669-1681. doi: 10.1007/s00247-025-06301-7. Epub 2025 Jun 17. Pediatr Radiol. 2025. PMID: 40526116
-
Noninvasive Assessment of Intracranial Pressure: Deformability Index as an Adjunct to Optic Nerve Sheath Diameter to Increase Diagnostic Ability.Neurocrit Care. 2024 Oct;41(2):479-488. doi: 10.1007/s12028-024-01955-x. Epub 2024 Mar 6. Neurocrit Care. 2024. PMID: 38448744 Free PMC article.
-
The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children. : Part II: age-related ONSD cut-off values and patency of the anterior fontanelle.Childs Nerv Syst. 2016 Oct;32(10):1779-85. doi: 10.1007/s00381-016-3068-4. Epub 2016 Sep 20. Childs Nerv Syst. 2016. PMID: 27659820
-
The relationship between transorbital ultrasound measurement of the optic nerve sheath diameter (ONSD) and invasively measured ICP in children : Part I: repeatability, observer variability and general analysis.Childs Nerv Syst. 2016 Oct;32(10):1769-78. doi: 10.1007/s00381-016-3067-5. Epub 2016 Sep 20. Childs Nerv Syst. 2016. PMID: 27659819
-
Using Optic Nerve Sheath Diameter for Intracranial Pressure (ICP) Monitoring in Traumatic Brain Injury: A Scoping Review.Neurocrit Care. 2024 Jun;40(3):1193-1212. doi: 10.1007/s12028-023-01884-1. Epub 2023 Dec 19. Neurocrit Care. 2024. PMID: 38114797 Free PMC article.
Cited by
-
An update on pediatric traumatic brain injury.Childs Nerv Syst. 2023 Nov;39(11):3071-3081. doi: 10.1007/s00381-023-06173-y. Epub 2023 Oct 6. Childs Nerv Syst. 2023. PMID: 37801113 Free PMC article. Review.
-
In Reply: Possible Factors Missed While Assessing Optic Nerve Sheath Diameter and Deformability Index in Patients with Severe Traumatic Brain Injury.Neurocrit Care. 2024 Aug;41(1):305-306. doi: 10.1007/s12028-024-02015-0. Epub 2024 Jun 20. Neurocrit Care. 2024. PMID: 38902579 No abstract available.
-
Diagnostic accuracy of ultrasonographic optic nerve sheath diameter compared to invasive ICP monitoring for intracranial hypertension in pediatric patients.Pediatr Radiol. 2025 Jul;55(8):1669-1681. doi: 10.1007/s00247-025-06301-7. Epub 2025 Jun 17. Pediatr Radiol. 2025. PMID: 40526116
-
Optic Nerve Ultrasound Evaluation in Children: A Review.Diagnostics (Basel). 2023 Feb 1;13(3):535. doi: 10.3390/diagnostics13030535. Diagnostics (Basel). 2023. PMID: 36766639 Free PMC article. Review.
-
Non-invasive assessment of ICP in children: advances in ultrasound-based techniques.Childs Nerv Syst. 2020 Jan;36(1):95-98. doi: 10.1007/s00381-019-04371-1. Epub 2019 Oct 4. Childs Nerv Syst. 2020. PMID: 31584121
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources