Comparative Study
. 2019 Jun;30(6):940-948.
doi: 10.1016/j.jvir.2018.05.005. Epub 2018 Aug 31.

Midterm Recanalization after Arterial Embolization Using Hydrogel-Coated Coils versus Fibered Coils in an Animal Model

Affiliations
Comparative Study

Midterm Recanalization after Arterial Embolization Using Hydrogel-Coated Coils versus Fibered Coils in an Animal Model

Audrey Fohlen et al. J Vasc Interv Radiol. 2019 Jun.

Abstract

Purpose: To compare angiographic and pathologic effects (ie, occlusion, recanalization) after embolization with Hydrogel-coated coils (HydroCoils) and fibered coils in the renal and internal iliac arteries after 7 days and 1 and 4 months in an animal model.

Materials and methods: Twelve sheep had 1 internal iliac and 1 renal artery randomly embolized with HydroCoils or fibered coils. Renal and internal iliac arteries were embolized with detachable 0.018-inch coils and pushable 0.035-inch coils, respectively. All animals had control angiography performed at 7 days, and 1 and 4 months to assess recanalization before euthanasia. Recanalization and inflammation were evaluated via pathologic examination.

Results: At 1 month, 100% of arteries embolized with HydroCoils were occluded vs 50% of those embolized with fibered coils (P = .004). At 4 months, 80% of arteries embolized with HydroCoils were occluded vs 25% of those embolized with fibered coils (P = .01). Surface of vessel occlusion was significantly greater for iliac arteries (96.7% ± 8.9) than for renal arteries (94.2% ± 5.3; P = .0076). Surface of occlusion of the renal arteries (92.2% ± 5.1) was lower for fibered coils than for HydroCoils (96.8% ± 4.7; P = .0287). Surface percentage of thrombus was significantly lower for HydroCoils than for fibered coils (P < .0001). Surface percentage of thrombus was correlated with surface percentage of recanalization (P = .0181).

Conclusions: After 4 months, 75% of arteries embolized with fibered coils were recanalized vs 20% of those embolized with HydroCoils (P = .01). Reduced amount of thrombus after embolization with HydroCoils accounted for a reduced rate of arterial recanalization.

PubMed Disclaimer

Publication types

LinkOut - more resources