. 2019 Aug;43(8):1891-1898.
doi: 10.1007/s00264-018-4136-z. Epub 2018 Sep 6.

Multiplex PCR Unyvero i60 ITI application improves detection of low-virulent microorganisms in periprosthetic joint infections

Affiliations

Multiplex PCR Unyvero i60 ITI application improves detection of low-virulent microorganisms in periprosthetic joint infections

Irene Katharina Sigmund et al. Int Orthop. 2019 Aug.

Abstract

Purpose: The aim of this study was to evaluate the pre-operative performance of an automated multiplex PCR (mPCR) system in patients with suspected periprosthetic joint infection (PJI).

Methods: Under sterile conditions, synovial fluid samples from patients with a suspected PJI were collected pre-operatively. One hundred eighty microliter of the aspirate was used for analysis in the mPCR. The remaining joint fluid was sent for microbiological analysis. PJI was diagnosed by using the Musculoskeletal Infection Society (MSIS) criteria. Total percentage agreement and Cohen's kappa coefficient were calculated to measure overall agreement.

Results: Overall, 90 patients with a suspected PJI were included. Using MSIS criteria, 38 (42%) patients were classified as septic. Total percent agreement between mPCR and synovial fluid culture was 86% with a Cohen's kappa of 0.68. The mPCR and synovial fluid culture showed sensitivities of 71% and 84%, respectively. Combined evaluation provided an even higher sensitivity of 92%. While Cutibacterium spp. were detected five times by mPCR, it could only be cultured once. A higher detection rate of CoNS by mPCR (n = 7) compared to conventional culture (n = 5) was also demonstrated. In comparison to synovial fluid culture, the mPCR missed Staphylococcus aureus five times.

Conclusion: With a moderate agreement between synovial fluid mPCR and culture, the mPCR system could be a useful adjunct in diagnosing a PJI pre-operatively. Due to faster availability of results and a higher detection rate of low-virulent microorganisms, it can complement conventional culture.

Keywords: Diagnosis; Low-grade infections; Low-virulent microorganisms; Multiplex PCR system; Periprosthetic joint infection.

PubMed Disclaimer

Conflict of interest statement

The authors declare that they have no conflict of interest.

Figures

Fig. 1
Fig. 1
Receiver operating characteristic curves for diagnostic accuracy of periprosthetic joint infection based on multiplex PCR (mPCR), synovial fluid culture (SF culture), and the combination of both test methods (combined)

Similar articles

Cited by

References

    1. Parvizi J, Gehrke T, Chen AF. Proceedings of the international consensus on periprosthetic joint infection. Bone Joint J. 2013;95-b(11):1450–1452. doi: 10.1302/0301-620X.95B11.33135. - DOI - PubMed
    1. Achermann Y, Vogt M, Leunig M, Wust J, Trampuz A. Improved diagnosis of periprosthetic joint infection by multiplex PCR of sonication fluid from removed implants. J Clin Microbiol. 2010;48(4):1208–1214. doi: 10.1128/JCM.00006-10. - DOI - PMC - PubMed
    1. Borde JP, Hacker GA, Guschl S, Serr A, Danner T, Hubner J, Burrack-Lange S, Ludke G, Helwig P, Hauschild O, Kern WV (2015) Diagnosis of prosthetic joint infections using UMD-Universal Kit and the automated multiplex-PCR Unyvero i60 ITI cartridge system: a pilot study. Infection. 10.1007/s15010-015-0796-4 - PubMed
    1. Hischebeth GT, Randau TM, Buhr JK, Wimmer MD, Hoerauf A, Molitor E, Bekeredjian-Ding I, Gravius S. Unyvero i60 implant and tissue infection (ITI) multiplex PCR system in diagnosing periprosthetic joint infection. J Microbiol Methods. 2016;121:27–32. doi: 10.1016/j.mimet.2015.12.010. - DOI - PubMed
    1. Portillo ME, Salvado M, Sorli L, Alier A, Martinez S, Trampuz A, Gomez J, Puig L, Horcajada JP. Multiplex PCR of sonication fluid accurately differentiates between prosthetic joint infection and aseptic failure. J Infect. 2012;65(6):541–548. doi: 10.1016/j.jinf.2012.08.018. - DOI - PubMed

LinkOut - more resources