Comparative Study
. 2018 Nov;211(5):1058-1062.
doi: 10.2214/AJR.18.19678. Epub 2018 Sep 12.

Radiation Exposure to the Sacroiliac Joint From Low-Dose CT Compared With Radiography

Affiliations
Comparative Study

Radiation Exposure to the Sacroiliac Joint From Low-Dose CT Compared With Radiography

Baljot S Chahal et al. AJR Am J Roentgenol. 2018 Nov.

Abstract

Objective: The primary objective of this study was to estimate the effective dose delivered to the sacroiliac joint (SIJ) from low-dose (LD) CT compared with that from radiography. Secondary objectives included evaluation of diagnostic quality of LD CT of the SIJ and development of a clinical protocol for LD CT of the SIJ.

Materials and methods: Data from 36 patients (19 women, 17 men) undergoing LD CT for suspected renal colic were analyzed. Two effective dose estimates were calculated: one for the SIJ and another for an extended region from the iliac crest to 1 cm below the SIJ. Thirty-six anteroposterior pelvic and 36 SIJ view radiographs were age-, sex-, and body width-matched to CT scans. Effective dose from radiography was estimated using the method described in International Commission on Radiologic Protection Publication 60.

Results: Maximum effective dose to the SIJ from LD CT was less than 1 mSv in all cases, with a mean ± SD of 0.42 ± 0.18 mSv (range, 0.14-0.83 mSv), whereas mean dose to the extended region was 0.57 ± 0.24 mSv (range, 0.19-1.11 mSv). Mean dose from SIJ radiographs was 0.15 ± 0.10 mSv (range, 0.07-1.38 mSv), and mean dose from a single pelvic radiograph was 0.09 ± 0.06 mSv (range, 0.04-0.37 mSv). All CT studies were of diagnostic quality for assessment of the SIJ.

Conclusion: LD CT of the SIJ can be consistently performed with an effective radiation dose of less than 1 mSv. Because reliability and sensitivity of radiography for sacroiliitis is poor, we recommend that LD CT replace radiography for dedicated evaluation of the SIJ.

Keywords: low-dose CT; radiation; radiography; sacroiliac joint; spondyloarthropathy.