. 2018 Dec:487:54-59.
doi: 10.1016/j.cca.2018.09.022. Epub 2018 Sep 12.

An automated method for measuring lipoprotein lipase and hepatic triglyceride lipase activities in post-heparin plasma

Affiliations

An automated method for measuring lipoprotein lipase and hepatic triglyceride lipase activities in post-heparin plasma

Kiyomi Nakajima et al. Clin Chim Acta. 2018 Dec.

Abstract

Background: Lipoprotein lipase (LPL) and hepatic triglyceride lipase (HTGL) play a central role in triglyceride-rich lipoprotein metabolism by catalyzing the hydrolysis of triglycerides. Quantification of LPL and HTGL activity is useful for diagnosing lipid disorders, but there has been no automated method for measuring these lipase activities.

Methods: The automated kinetic colorimetric method was used for assaying LPL and HTGL activity in the post-heparin plasma using the natural long-chain fatty acid 2-diglyceride as a substrate. LPL activity was determined with apoCII and HTGL activity was determined without apoCII with 2 channel of auto-analyzer.

Results: The calibration curve for dilution tests of the LPL and HTGL activity assay ranged from 0.0 to 500 U/L. Within-run CV was obtained within a range of 5%. No interference was observed in the testing of specimens containing potentially interfering substances. The measurement range of LPL activity in the post-heparin plasma was 30-153 U/L, while HTGL activity was 135-431 U/L in normal controls.

Conclusions: The L PL and HTGL activity assays are applicable to quantitating the LPL and HTGL activity in the post-heparin plasma. This assay is more convenient and faster than radiochemical assay and highly suitable for the detection of lipid disorders.

Keywords: Automated lipase activity assay; Cardiac catheter patients; Hepatic triglyceride lipase (HTGL); Lipoprotein lipase (LPL); Post-heparin plasma; apoCII.