Evaluation of quantitative 123I and 131I SPECT with Monte Carlo-based down-scatter compensation
- PMID: 30222722
- DOI: 10.1097/MNM.0000000000000920
Evaluation of quantitative 123I and 131I SPECT with Monte Carlo-based down-scatter compensation
Abstract
Objective: Quantitative I and I single-photon emission computed tomography (SPECT) is hampered by down-scatter from the high-energy peaks. This paper presents a down-scatter compensation method, where down-scatter generated in the patient and gamma camera collimator and detector is modelled using Monte Carlo simulation in the ordered subsets expectation maximization SPECT reconstruction algorithm.
Materials and methods: The new down-scatter compensation method was compared with conventional triple energy window (TEW) scatter compensation and Gaussian convolution-based forced detection Monte Carlo methods. The comparison was made with the NEMA-IEC phantom using six spherical inserts (diameters from 10 to 37 mm) and a lung compartment. The phantom was filled with I and I solutions to known sphere-to-background concentration ratios. Spherical volumes of interest with the same diameter as the inserts were drawn on the images, and recovery coefficients for the spheres were calculated in addition to lung-to-background ratio.
Results: The new down-scatter compensation method provided higher recovery coefficients than the TEW scatter compensation or Gaussian convolution-based forced detection Monte Carlo algorithm for both isotopes. Background activity concentration could be accurately estimated with the new down-scatter compensation method and with the TEW scatter compensation, whereas activity concentration of the spheres was severely underestimated even with the new method.
Conclusion: Down-scatter compensation with Monte Carlo-simulation effectively reduces down-scatter effects in I and I SPECT imaging.
Similar articles
-
Impact of reconstruction parameters on quantitative I-131 SPECT.Phys Med Biol. 2016 Jul 21;61(14):5166-82. doi: 10.1088/0031-9155/61/14/5166. Epub 2016 Jun 28. Phys Med Biol. 2016. PMID: 27352225
-
Quantitative Monte Carlo-based holmium-166 SPECT reconstruction.Med Phys. 2013 Nov;40(11):112502. doi: 10.1118/1.4823788. Med Phys. 2013. PMID: 24320461
-
Scatter and crosstalk corrections for (99m)Tc/(123)I dual-radionuclide imaging using a CZT SPECT system with pinhole collimators.Med Phys. 2015 Dec;42(12):6895-911. doi: 10.1118/1.4934830. Med Phys. 2015. PMID: 26632046
-
Application of Monte Carlo Algorithms to Cardiac Imaging Reconstruction.Curr Pharm Des. 2021;27(16):1960-1972. doi: 10.2174/1381612826999201228215225. Curr Pharm Des. 2021. PMID: 33371829 Review.
-
Quantitative myocardial perfusion SPECT.J Nucl Cardiol. 1998 Sep-Oct;5(5):507-22. doi: 10.1016/s1071-3581(98)90182-9. J Nucl Cardiol. 1998. PMID: 9796898 Review.
Cited by
-
Optimization of SPECT/CT imaging protocols for quantitative and qualitative 99mTc SPECT.EJNMMI Phys. 2021 Jul 30;8(1):57. doi: 10.1186/s40658-021-00405-3. EJNMMI Phys. 2021. PMID: 34328565 Free PMC article.
-
Evaluation of reconstruction methods and image noise levels concerning visual assessment of simulated liver lesions in 111In-octreotide SPECT imaging.EJNMMI Phys. 2023 Jun 2;10(1):36. doi: 10.1186/s40658-023-00557-4. EJNMMI Phys. 2023. PMID: 37266738 Free PMC article.
-
Multiplexed imaging of radionuclides.Nat Biomed Eng. 2025 Jul;9(7):993-1006. doi: 10.1038/s41551-025-01406-8. Epub 2025 Jun 20. Nat Biomed Eng. 2025. PMID: 40542108 Free PMC article. Review.
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Other Literature Sources