Loop and knot security of a novel arthroscopic sliding-locking knot using high-strength sutures
- PMID: 30224854
- PMCID: PMC6138853
- DOI: 10.1016/j.jor.2018.08.031
Loop and knot security of a novel arthroscopic sliding-locking knot using high-strength sutures
Erratum in
-
Erratum regarding missing Declaration of Competing Interest statements in previously published articles.J Orthop. 2020 Dec 14;23:273. doi: 10.1016/j.jor.2020.12.001. eCollection 2021 Jan-Feb. J Orthop. 2020. PMID: 33746417 Free PMC article.
Abstract
Objective: We evaluated the loop and knot security of a novel arthroscopic knot, the Wiese knot, using different types of sutures.
Methods: The Wiese knot was tied using four different brands of braided sutures (Ethibond, Orthocord, FiberWire, and UltraBraid) with and without a series of three reversing half-hitches (RHAPs) and tested for loop and knot security.
Results: Orthocord provided the greatest amount of loop security. FiberWire delivered the highest knot security. UltraBraid had the greatest ultimate force. Three half-hitches increased the maximal load to clinical failure.
Conclusion: The biomechanical characteristics of the Wiese knot are affected by suture material qualities.
Keywords: Arthroscopic knots; Biomechanical strength.
Figures
References
-
- Burkhart S.S., Wirth M.A., Simonich M., Salem D., Lanctot D., Athanasiou K. Knot security in simple sliding knots and its relationship to rotator cuff repair: how secure must the knot be? Arthroscopy. 2000;16(2):202–207. - PubMed
-
- Abbi G., Espinoza L., Odell T., Mahar A., Pedowitz R. Evaluation of 5 knots and 2 suture materials for arthroscopic rotator cuff repair: very strong sutures can still slip. Arthroscopy. 2006;22(1):38–43. - PubMed
-
- Barber F.A., Herbert M.A., Beavis R.C. Cyclic load and failure behavior of arthroscopic knots and high strength sutures. Arthroscopy. 2009;25(2):192–199. - PubMed
-
- Dahl K.A., Patton D.J., Dai Q., Wongworawat M.D. Biomechanical characteristics of 9 arthroscopic knots. Arthroscopy. 2010;26(6):813–818. - PubMed
-
- Hassinger S.M., Wongworawat M.D., Hechanova J.W. Biomechanical characteristics of 10 arthroscopic knots. Arthroscopy. 2006;22(8):827–832. - PubMed
LinkOut - more resources