Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Case Reports
. 2018 Jun 15;4(9):408-411.
doi: 10.1016/j.hrcr.2018.06.004. eCollection 2018 Sep.

Inappropriate shock from delayed T-wave oversensing by a subcutaneous implantable cardioverter-defibrillator after septal myectomy for hypertrophic cardiomyopathy

Affiliations
Case Reports

Inappropriate shock from delayed T-wave oversensing by a subcutaneous implantable cardioverter-defibrillator after septal myectomy for hypertrophic cardiomyopathy

Benjamin C Salgado et al. HeartRhythm Case Rep. .
No abstract available

Keywords: Hypertrophic cardiomyopathy; Left bundle branch block; SICD; Septal myectomy; T-wave oversensing.

PubMed Disclaimer

Figures

Figure 1
Figure 1
A: A 12-lead electrocardiogram prior to septal myectomy showing sinus rhythm, left atrial enlargement, and left axis deviation with a QRS duration of 88 ms. B: A 12-lead electrocardiogram soon after septal myectomy demonstrating normal sinus rhythm, new left bundle branch block, and a QRS duration of 160 ms.
Figure 2
Figure 2
A 12-lead electrocardiogram (ECG) 39 days after septal myectomy showing sinus rhythm with left bundle branch block and a QRS duration of 154 ms. Notice the taller T waves compared to the ECG done shortly after myectomy.
Figure 3
Figure 3
Subcutaneous implantable cardioverter-defibrillator interrogation strip showing sinus tachycardia and inappropriate shock from T-wave oversensing.
Supplemental Figure
Supplemental Figure
Panels A–C show S-ECG tracings comparing primary vector pre- and post-myectomy as well as the alternate vector following the ICD shocks. A: S-ECG at baseline (Pre-Myectomy). B: S-ECG 3 months after myectomy. C: S-ECG 2 months post-shock (sensing vector is alternate with SMARTPASS on). Note QRS axis compared to the T-wave.
Supplemental Figure
Supplemental Figure
Panels A–C show S-ECG tracings comparing primary vector pre- and post-myectomy as well as the alternate vector following the ICD shocks. A: S-ECG at baseline (Pre-Myectomy). B: S-ECG 3 months after myectomy. C: S-ECG 2 months post-shock (sensing vector is alternate with SMARTPASS on). Note QRS axis compared to the T-wave.
Supplemental Figure
Supplemental Figure
Panels A–C show S-ECG tracings comparing primary vector pre- and post-myectomy as well as the alternate vector following the ICD shocks. A: S-ECG at baseline (Pre-Myectomy). B: S-ECG 3 months after myectomy. C: S-ECG 2 months post-shock (sensing vector is alternate with SMARTPASS on). Note QRS axis compared to the T-wave.

References

    1. Bardy G.H., Smith W.M., Hood M.A. An entirely subcutaneous implantable cardioverter-defibrillator. N Engl J Med. 2010;363:36–44. - PubMed
    1. Bellardine Black C.L., Stromberg K., van Balen G.P., Ghanem R.N., Breedveld R.W., Tieleman R.G. Is surface ECG a useful surrogate for subcutaneous ECG? Pacing Clin Electrophysiol. 2010;33:135–145. - PubMed
    1. Francia P., Adduci C., Palano F., Semprini L., Serdoz A., Montesanti D., Santini D., Musumeci B., Salvati A., Volpe M., Autore C. Eligibility for the subcutaneous implantable cardioverter-defibrillator in patients with hypertrophic cardiomyopathy. J Cardiovasc Electrophysiol. 2015;26:893–899. - PubMed
    1. Randles D.A., Hawkins N.M., Shaw M., Patwala A.Y., Pettit S.J., Wright D.J. How many patients fulfil the surface electrocardiogram criteria for subcutaneous implantable cardioverter-defibrillator implantation? Europace. 2014;16:1015–1021. - PubMed
    1. Olde Nordkamp L.R.A., Warnaars J.L.F., Kooiman K.M., de Groot J.R., Rosenmöller B.R.A.M., Wilde A.A.M., Knops R.E. Which patients are not suitable for a subcutaneous ICD: incidence and predictors of failed QRS-T-wave morphology screening. J Cardiovasc Electrophysiol. 2014;25:494–499. - PubMed

Publication types

LinkOut - more resources