Single and dual coil shock efficacy and predictors of shock failure in patients with modern implantable cardioverter defibrillators-a single-center paired randomized study
- PMID: 30229405
- DOI: 10.1007/s10840-018-0443-y
Single and dual coil shock efficacy and predictors of shock failure in patients with modern implantable cardioverter defibrillators-a single-center paired randomized study
Abstract
Purpose: Implantable cardioverter defibrillators (ICDs) can treat life-threatening tachyarrhythmia with high-voltage shocks. The aims were to compare the efficacy of single and dual coil shock vectors in modern ICDs and to identify predictors of shock failure.
Methods: This is a single-center paired randomized study including 216 patients with mixed indications and ICDs from four manufacturers. All patients underwent two implant defibrillation tests using single and dual coil vectors with the test order randomized. Tested shock energy differed slightly between manufacturers because of differences in device programmability: first shock approximately 15 J below maximal output-if failed, second shock approximately 10 J below maximal output-if failed, third shock at maximal output.
Results: First shock success rate was 399/432 (92.4%). Comparing single and dual coil vectors, no differences were seen in first shock efficacy (91.7% vs. 93.1%, P = 0.629) or lowest tested succesfully stored energy (27.2 J vs. 27.1 J, P = 0.620). All successive internal shocks failed in 4/432 (0.9%) of inductions requiring external rescue shocks to restore circulation. Multivariate predictors of first shock failure were QRS duration (relative risk 0.81 per 10 ms, P = 0.001), amiodarone treatment (relative risk 3.30, P = 0.003), and body height (relative risk 1.70 per 10 cm, P = 0.019).
Conclusions: Implant defibrillation testing of modern intravenous ICD systems demonstrates high shock efficacy with no difference between single and dual coil vectors.
Keywords: Coil; Defibrillation; Implantable cardioverter defibrillator; Leads; Shock efficacy.
Similar articles
-
Multicentre comparison Of shock efficacy using single-vs. Dual-coil lead systems and Anodal vs. cathodaL polarITY defibrillation in patients undergoing transvenous cardioverter-defibrillator implantation. The MODALITY study.J Interv Card Electrophysiol. 2015 Jun;43(1):45-54. doi: 10.1007/s10840-015-9980-9. Epub 2015 Feb 19. J Interv Card Electrophysiol. 2015. PMID: 25690336 Clinical Trial.
-
Inductionless or limited shock testing is possible in most patients with implantable cardioverter- defibrillators/cardiac resynchronization therapy defibrillators: results of the multicenter ASSURE Study (Arrhythmia Single Shock Defibrillation Threshold Testing Versus Upper Limit of Vulnerability: Risk Reduction Evaluation With Implantable Cardioverter-Defibrillator Implantations).Circulation. 2007 May 8;115(18):2382-9. doi: 10.1161/CIRCULATIONAHA.106.663112. Epub 2007 Apr 30. Circulation. 2007. PMID: 17470697 Clinical Trial.
-
Single-coil and dual-coil defibrillator leads and association with clinical outcomes in a complete Danish nationwide ICD cohort.Heart Rhythm. 2016 Mar;13(3):706-12. doi: 10.1016/j.hrthm.2015.11.034. Epub 2015 Nov 22. Heart Rhythm. 2016. PMID: 26593333
-
Outcomes with single-coil versus dual-coil implantable cardioverter defibrillators: a meta-analysis.Europace. 2018 Mar 1;20(3):e21-e29. doi: 10.1093/europace/euw438. Europace. 2018. PMID: 28339860 Review.
-
The dilemma of ICD implant testing.Pacing Clin Electrophysiol. 2007 May;30(5):675-700. doi: 10.1111/j.1540-8159.2007.00730.x. Pacing Clin Electrophysiol. 2007. PMID: 17461879 Review.
Cited by
-
Comparison of Single-Coil Versus Dual-Coil Implantable Cardioverter Defibrillator Devices: A Systematic Review and Meta-Analysis of Efficacy and Extraction-Related Outcomes.Clin Cardiol. 2025 Feb;48(2):e70083. doi: 10.1002/clc.70083. Clin Cardiol. 2025. PMID: 39910823 Free PMC article.
-
Comparative efficacy of single-coil versus dual-coil ICD leads: a meta-analysis of clinical outcomes.Future Cardiol. 2025 Mar;21(3):167-175. doi: 10.1080/14796678.2025.2459542. Epub 2025 Jan 30. Future Cardiol. 2025. PMID: 39882757
References
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical