The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth
- PMID: 30252806
- DOI: 10.1097/PRS.0000000000004801
The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth
Abstract
Background: The introduction of texture to the outer shell of breast implants was aimed at increasing tissue incorporation and reducing capsular contracture. It has also been shown that textured surfaces promote a higher growth of bacteria and are linked to the development of breast implant-associated anaplastic large cell lymphoma.
Methods: The authors aimed to measure the surface area and surface roughness of 11 available implants. In addition, the authors aimed to subject these implant shells to an in vitro bacterial attachment assay with four bacterial pathogens (Staphylococcus epidermidis, S. aureus, Pseudomonas aeruginosa, and Ralstonia pickettii) and study the relationship among surface area, surface roughness, and bacterial growth.
Results: Surface area measurement showed grouping of implants into high, intermediate, low, and minimal. Surface roughness showed a correlation with surface area. The in vitro assay showed a significant linear relationship between surface area and bacterial attachment/growth. The high surface area/roughness implant texture grew significantly more bacteria at 24 hours, whereas the minimal surface area/roughness implant textures grew significantly fewer bacteria of all types at 24 hours. For implants with intermediate and low surface areas, some species differences were observed, indicating possible affinity of specific bacterial species to surface morphology.
Conclusions: Implant shells should be reclassified using surface area/roughness into four categories (high, intermediate, low, and minimal). This classification is superior to the use of descriptive terms such as macrotexture, microtexture, and nanotexture, which are not well correlated with objective measurement and/or functional outcomes.
Comment in
-
The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth.Plast Reconstr Surg. 2019 Jul;144(1):143e-144e. doi: 10.1097/PRS.0000000000005732. Plast Reconstr Surg. 2019. PMID: 31246849 No abstract available.
-
Reply: The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth.Plast Reconstr Surg. 2019 Jul;144(1):144e. doi: 10.1097/PRS.0000000000005733. Plast Reconstr Surg. 2019. PMID: 31246850 No abstract available.
-
Rough Textured Silicone Implants, Bacterial Biofilms, and Capsular Contracture.Plast Reconstr Surg. 2019 Aug;144(2):322e-323e. doi: 10.1097/PRS.0000000000005816. Plast Reconstr Surg. 2019. PMID: 31348383 No abstract available.
-
Reply: The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth.Plast Reconstr Surg. 2019 Aug;144(2):323e-324e. doi: 10.1097/PRS.0000000000005818. Plast Reconstr Surg. 2019. PMID: 31348384 No abstract available.
-
The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth.Plast Reconstr Surg. 2019 Nov;144(5):929e-930e. doi: 10.1097/PRS.0000000000006109. Plast Reconstr Surg. 2019. PMID: 31403569 No abstract available.
-
Reply: The Functional Influence of Breast Implant Outer Shell Morphology on Bacterial Attachment and Growth.Plast Reconstr Surg. 2019 Nov;144(5):930e-931e. doi: 10.1097/PRS.0000000000006110. Plast Reconstr Surg. 2019. PMID: 31688774 No abstract available.
References
-
- Ashley FL. Further studies on the natural-Y breast prosthesis. Plast Reconstr Surg. 1972;49:414419.
-
- Chan SC, Birdsell DC, Gradeen CY. Urinary excretion of free toluenediamines in a patient with polyurethane-covered breast implants. Clin Chem. 1991;37:21432145.
-
- Chan SC, Birdsell DC, Gradeen CY. Detection of toluenediamines in the urine of a patient with polyurethane-covered breast implants. Clin Chem. 1991;37:756758.
-
- Henderson PW, Nash D, Laskowski M, Grant RT. Objective comparison of commercially available breast implant devices. Aesthetic Plast Surg. 2015;39:724732.
-
- Sforza M, Zaccheddu R, Alleruzzo A, et al. Preliminary 3-year evaluation of experience with SilkSurface and VelvetSurface Motiva silicone breast implants: A single-center experience with 5813 consecutive breast augmentation cases. Aesthet Surg J. 2018;38(Suppl 2):S62S73.
MeSH terms
LinkOut - more resources
Full Text Sources
Other Literature Sources
Medical