Skip to main page content
U.S. flag

An official website of the United States government

Dot gov

The .gov means it’s official.
Federal government websites often end in .gov or .mil. Before sharing sensitive information, make sure you’re on a federal government site.

Https

The site is secure.
The https:// ensures that you are connecting to the official website and that any information you provide is encrypted and transmitted securely.

Access keys NCBI Homepage MyNCBI Homepage Main Content Main Navigation
Comparative Study
. 2019 Mar;54(2):151-159.
doi: 10.1007/s10840-018-0451-y. Epub 2018 Sep 25.

Efficacy and safety of non-transvenous cardioverter defibrillators in infants and young children

Affiliations
Comparative Study

Efficacy and safety of non-transvenous cardioverter defibrillators in infants and young children

Matthias J Müller et al. J Interv Card Electrophysiol. 2019 Mar.

Abstract

Background: Implantable cardioverter defibrillators (ICD) protect from sudden cardiac death (SCD). In infants and young children, ICD implantation and programming is challenging due to small body size, elevated heart rates, and high physical activity.

Purpose: We report our experience applying a non-transvenous ICD (NT-ICD) system to infants and children < 12 years of age and < 45-kg body weight.

Methods: Between 07/2004 and 07/2016, NT-ICD had been implanted in 36 patients. Nine out of 36 patients (25%) had NT-ICD implantation for primary and 27/36 (75%) for secondary prevention. Underlying diseases included inherited primary electrical arrhythmogenic diseases (n = 26; 72%), cardiomyopathies (n = 8; 22%), and congenital heart defects (n = 2; 6%). The median (interquartile range) age at implantation was 6 (1.9-8.4) years, and the median body weight was 21.7 (11.2-26.8) kg. Three different NT-ICD implantation techniques had been applied over time: (1) abdominal device/subcutaneous shock coil, (2) abdominal device/pleural shock coil, and (3) subcardiac device/pleural shock coil.

Results: During median follow-up of 5.2 (2.7-7.2) years, appropriate ICD discharges were documented in 12 (33.3%) and inappropriate shocks in 4 patients (11.1%). In 12/36 individuals (33.3%), a total of 25 surgical revisions were required due to NT-ICD malfunction. Eighteen out of 25 (72%) surgical revisions were necessary in patients with subcutaneous shock coil/abdominal device position. Surgical revisions (3/25, 12%) were significantly reduced (p < 0.001) after modifying the implantation technique to subcardiac device/pleural shock coil.

Conclusions: NT-ICD was safe and effective in infants and young children. Appropriate ICD discharges occurred in a considerable number of patients. After modifying the implantation technique, the need for surgical revision could significantly be decreased.

Keywords: Children; Defibrillation threshold; ICD discharge; Infants; Non-transvenous cardioverter defibrillator.

PubMed Disclaimer

References

    1. Pacing Clin Electrophysiol. 2004 Jul;27(7):924-32 - PubMed
    1. J Cardiovasc Electrophysiol. 2005 Jan;16(1):69-73 - PubMed
    1. Z Kardiol. 2005 Jun;94(6):415-8 - PubMed
    1. J Cardiovasc Electrophysiol. 2006 Jan;17(1):41-6 - PubMed
    1. J Cardiovasc Electrophysiol. 2006 Jul;17(7):754-9 - PubMed

Publication types

MeSH terms

LinkOut - more resources