Observational Study
. 2019 Mar;33(3):404-410.
doi: 10.1038/s41433-018-0221-1. Epub 2018 Sep 28.

Functional assessment of a new extended depth-of-focus intraocular lens

Affiliations
Observational Study

Functional assessment of a new extended depth-of-focus intraocular lens

Giacomo Savini et al. Eye (Lond). 2019 Mar.

Abstract

Background/objective: This study aimed to investigate the visual performance of a new extended depth-of-focus intraocular lens (EDOF-IOL).

Subjects/methods: In this multicenter, prospective, observational study, we enrolled 97 patients who underwent cataract surgery or refractive lens exchange with implantation of the Mini Well EDOF-IOL (SIFI, Italy). Patients underwent postoperatively the following examinations between 4 and 8 weeks after surgery: corrected distance visual acuity (CDVA), reading speed with Radner's chart, distance-corrected near visual acuity (DCNVA), defocus curve, contrast sensitivity, and haloes quantitative assessment.

Results: In the whole sample, the mean monocular CDVA and DCNVA were, respectively, 0.02 ± 0.07 logMAR and 0.38 ± 0.15 logRAD (logarithm of the reading acuity determination). In the 67 bilaterally implanted patients, binocular CDVA and DCNVA were better (0.00 ± 0.05 logMAR and 0.26 ± 0.13 logRAD) than the corresponding monocular values (p = 0.02 and p = 0.0002, respectively). Ninety-two percent of patients bilaterally implanted reached a binocular reading speed >80 words per minute at a 0.5 logRAD print size (corresponding to the common book print size). The defocus curves showed that the EDOF-IOL provided increased depth of focus through 2.0 D of defocus, with the best performance at 1.0 and 1.5 D. Contrast sensitivity was within normal limits at all spatial frequencies. The mean visual disturbance index was 0.08 ± 0.12, suggesting low night visual disturbances.

Conclusions: The new EDOF-IOL provided good visual acuity for distance, intermediate, and near vision, with no loss of contrast sensitivity and low risk of night visual disturbances.

PubMed Disclaimer

Conflict of interest statement

GS and RB are consultants of SIFI.

Figures

Fig. 1
Fig. 1
Mean values of monocular and binocular reading speed obtained at different print sizes with the Mini Well EDOF-IOL. WPM words per minute, logRAD logarithm of reading acuity determination
Fig. 2
Fig. 2
Monocular defocus curve in patients implanted with the Mini Well EDOF-IOL. The vertical bars represent the standard deviation at defocus. CDVA corrected distance visual acuity
Fig. 3
Fig. 3
Contrast sensitivity of the Mini Well EDOF-IOL. The dotted lines represent the minimum (min) and maximum (max) limits of the normative database. The vertical bars represent the standard deviation at each frequency

References

    1. Savini G, Hoffer KJ, Lombardo M, Serrao S, Schiano-Lomoriello D, Ducoli P. Influence of the effective lens position, as predicted by axial length and keratometry, on the near add power of multifocal intraocular lenses. J Cataract Refract Surg. 2016;42:44–49. doi: 10.1016/j.jcrs.2015.07.044. - DOI - PubMed
    1. Alfonso JF, Fernández-VegaL, Baamonde MB, Montés-Micó R. Prospective visual evaluation of apodized diffractive intraocular lenses. J Cataract Refract Surg. 2007;33:1235–43. doi: 10.1016/j.jcrs.2007.03.034. - DOI - PubMed
    1. Kohnen T, Nuijts R, Levy P, Haefliger E, Alfonso JF. Visual function after bilater implantation of apodized diffractive aspheric multifocal intraocular lenses with a +3.0 D addition. J Cataract Refract Surg. 2009;35:2062–9. doi: 10.1016/j.jcrs.2009.08.013. - DOI - PubMed
    1. Maxwell WA, Cionni RJ, Lehmann RP, Modi SS. Functional outcomes after bilateral implantation of apodized diffractive aspheric acrylic intraocular lenses with a +3.0 or +4.0 diopter addition power. J Cataract Refract Surg. 2009;35:2054–61. doi: 10.1016/j.jcrs.2009.06.041. - DOI - PubMed
    1. Pedrotti E, Mastropasqua R, Passilongo M, Parisi G, Marchesoni I, Marchini G. Comparison of two multifocal intraocular lens designs that differ only in near add. J Refract Surg. 2014;30:754–60. doi: 10.3928/1081597X-20141021-07. - DOI - PubMed

Publication types

LinkOut - more resources