Assessment of isocenter alignment during CT colonography: Implications for clinical practice
- PMID: 30292502
- DOI: 10.1016/j.radi.2018.04.003
Assessment of isocenter alignment during CT colonography: Implications for clinical practice
Abstract
Introduction: Optimization of image quality and patient radiation dose is achieved in part by positioning the patient at the isocenter of the CT gantry. The aim of this study was to establish whether there was increased isocenter misalignment (IM) in CT colonography (CTC) scans by comparing patient position during the prone part of a CTC to patient position during renal stone protocol CT (CT-KUB) and patient position during the supine part of a CTC to patient position during abdominopelvic CT (CT-AP).
Methods: Two hundred and twenty two consecutive outpatient adult CTC studies performed between January and December 2016 were retrospectively analyzed. Automated dose-tracking software was used to quantify IM in the x and y planes. Renal stone CT-KUB (n = 100) and standard CT-AP (n = 100) were used as comparison studies.
Results: IM during CTC was significantly greater in the y-axis compared with the x-axis for both prone (p = 0.002) and supine (p < 0.001) scanning. IM was significantly greater during prone CTC compared with CT-KUB (p = 0.008) and during supine CTC compared with CT-AP (p = 0.0001). IM was shown to be slightly greater in studies performed by more experienced radiographers (p = 0.04). IM was not associated with patient age, gender or size (p > 0.05 for all).
Conclusion: Isocenter misalignment is greater during CT colonography compared with CT-KUB or CT-AP. Strategies for improving patient positioning could include radiographer education and automated patient centering solutions.
Keywords: Adult; Colonography, Computed tomographic; Patient positioning; Tomography, X-ray computed.
Copyright © 2018 The College of Radiographers. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Volumetric analysis of colonic distention according to patient position at CT colonography: diagnostic value of the right lateral decubitus series.AJR Am J Roentgenol. 2014 Dec;203(6):W623-8. doi: 10.2214/AJR.13.12369. AJR Am J Roentgenol. 2014. PMID: 25415727 Free PMC article.
-
Patient centring and scan length: how inaccurate practice impacts on radiation dose in CT colonography (CTC).Radiol Med. 2019 Aug;124(8):762-767. doi: 10.1007/s11547-019-01021-z. Epub 2019 Mar 8. Radiol Med. 2019. PMID: 30848421
-
Ultra-low dose CT colonography with automatic tube current modulation and sinogram-affirmed iterative reconstruction: Effects on radiation exposure and image quality.J Appl Clin Med Phys. 2019 Jan;20(1):321-330. doi: 10.1002/acm2.12510. Epub 2018 Dec 26. J Appl Clin Med Phys. 2019. PMID: 30586479 Free PMC article.
-
Computed tomographic colonography (CTC): Possibilities and limitations of clinical application in colorectal polyps and cancer.Technol Cancer Res Treat. 2004 Apr;3(2):201-7. doi: 10.1177/153303460400300213. Technol Cancer Res Treat. 2004. PMID: 15059026 Review.
-
Multislice CT colonography: technical developments.Semin Ultrasound CT MR. 2001 Oct;22(5):425-31. doi: 10.1016/s0887-2171(01)90035-6. Semin Ultrasound CT MR. 2001. PMID: 11665921 Review.
Cited by
-
Effects of combining multiple dose reduction techniques on coronary computed tomography angiography.World J Radiol. 2023 Feb 28;15(2):32-41. doi: 10.4329/wjr.v15.i2.32. World J Radiol. 2023. PMID: 36874259 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources