Review
. 2018 Nov;66(11):827-833.
doi: 10.1007/s00106-018-0568-4.

[Reconstructive orbital surgery]

[Article in German]
Affiliations
Review

[Reconstructive orbital surgery]

[Article in German]
M E H Wagner et al. HNO. 2018 Nov.

Abstract

Background: In complex orbital reconstructions ideal positioning of the bony buttress and surrounding soft tissue is a prerequisite for an aesthetic and functional result. The use of computer-assisted surgery can support the surgeon before and during the reconstruction procedure and facilitate quality control processes. This is illustrated using three clinical examples.

Material and methods: The areas of interest (bony defect areas, surrounding tissues) are segmented in a 3D image series. In most cases, the contralateral non-affected side can serve as the reference in virtual reconstruction. The virtual model obtained can now be used for the manufacturing of patient specific models and implants, as well as for intraoperative navigation or direct quality control with the use of intraoperative cone beam computed tomography (CBCT).

Results: For the reconstruction of primary and secondary traumatic defects as well as for congenital malformations or neoplastic diseases, the presented workflow can be used. Preoperative virtual visualization, patient specific reconstruction and direct quality control using intraoperative CBCT ensure that the preoperatively planned result can be achieved. Together with the interplay of hard and soft tissue the best possible results can be achieved.

Conclusion: Computer-assisted surgery has been continuously further developed over the last two decades and is currently used in the clinical routine. Patient specific implants in combination with the use of direct intraoperative quality control facilitate the reconstruction of complex orbital injuries and defects and enable the ideal reconstruction from both aesthetic and functional aspects.

Keywords: Computer-assisted surgery; Intraoperative X-rays; Patient-specific implant; Quality control; Titanium.

PubMed Disclaimer

References

    1. J Craniomaxillofac Surg. 2015 Jan;43(1):126-30 - PubMed
    1. J Craniomaxillofac Surg. 1998 Feb;26(1):11-6 - PubMed
    1. Radiat Oncol. 2011 Nov 16;6:159 - PubMed
    1. J Craniomaxillofac Surg. 2015 Oct;43(8):1340-7 - PubMed
    1. J Craniomaxillofac Surg. 2016 Sep;44(9):1485-97 - PubMed

LinkOut - more resources