Role of a mydriasis-free, full-field flicker ERG device in the detection of diabetic retinopathy

Affiliations

Role of a mydriasis-free, full-field flicker ERG device in the detection of diabetic retinopathy

Mehmet Fatih Kağan Değirmenci et al. Doc Ophthalmol. 2018 Dec.

Abstract

Purpose: To determine if the RETeval system can be used for the screening of diabetic retinopathy (DR) to provide early diagnosis.

Methods: The subjects were 42 diabetic patients selectively recruited by examination of their medical records to have varying severities of DR. The severity of DR was classified into four groups according to the International Clinical Diabetic Retinopathy Disease Severity Scale. Full-field electroretinograms (ERG) without mydriasis were obtained by the DR assessment protocol of the RETeval system. Macular retinal nerve fiber layer (RNFL) thickness was measured by optical coherence tomography. We compared the DR assessment protocol results and the macular RNFL thickness among four groups. Moreover, an analysis was conducted on whether there was any correlation among the DR assessment protocol results, duration of diabetes mellitus, and RNFL thickness individually for each group of patients.

Results: The mean ages and mean duration of diabetes mellitus of the four groups were similar. The DR assessment protocol results in the moderate-severe nonproliferative DR, and proliferative DR groups were significantly higher than those in the other groups (p < 0.001). The mean macular RNFL thickness was similar in all groups. No significant correlation was found between the DR assessment protocol results and duration of DM and the RNFL thickness.

Conclusions: Our results suggest that the RETeval full-field ERG system can be used as an adjunctive tool for the mass screening of DR, while macular RNFL thickness would not be useful.

Keywords: Diabetic retinopathy; Electroretinogram (ERG); Flicker ERG; RETeval; Screening.