The effect of radial head prosthesis diameter on posterolateral rotatory instability of the elbow

Affiliations

The effect of radial head prosthesis diameter on posterolateral rotatory instability of the elbow

Chad E Songy et al. Clin Biomech (Bristol). 2018 Dec.

Abstract

Background: The purpose of this study is to investigate how different diameters of radial head replacement affect posterolateral translation with a valgus and supination force. We hypothesized that there would be less posterolateral rotatory translation with larger implant diameter.

Methods: Eleven cadaveric arms were stressed at 30 and 60° of flexion with a consistent supination and valgus stress force under five conditions: native radial head, radial head excision, and with 3 sizes of radial head prosthesis. Displacement of the radial head posteriorly in relation to the capitellum on radiographs was measured. Displacement was expressed as a percentage relative to the average of the maximum and minimum native radial head diameters.

Findings: The native radial heads had average minimum and maximum diameters of 23.3 mm and 25.2 mm, respectively. The angle of testing did not significantly change translation of the radial head. There was increased posterior translation relative to native head as the radial head sizes decreased from 24 mm to 20 mm and with excision of the radial head. Compared to the native head, the differences in displacement were statistically significant for the 20 mm radial head, but not for the 22 mm or 24 mm replacements. Radial head translation significantly increased after radial head excision.

Interpretations: This cadaveric study illustrates that patients treated with radial head excision and radial head prosthesis with undersized diameters have increased posterior translation with a valgus and supination stress. The larger the radial head prosthesis (closer to native radial head), the more closely it approximated the amount of translation of the native radial head.

Keywords: Elbow; Lateral collateral ligament; Radial head replacement; Rotatory posterolateral instability.