Error detection during VMAT delivery using EPID-based 3D transit dosimetry
- PMID: 30337003
- DOI: 10.1016/j.ejmp.2018.10.005
Error detection during VMAT delivery using EPID-based 3D transit dosimetry
Abstract
Purpose: To investigate the effectiveness of an EPID-based 3D transit dosimetry system in detecting deliberately introduced errors during VMAT delivery.
Methods: An Alderson phantom was irradiated using four VMAT treatment plans (one prostate, two head-and-neck and one lung case) in which delivery, thickness and setup errors were introduced. EPID measurements were performed to reconstruct 3D dose distributions of "error" plans, which were compared with "no-error" plans using the mean gamma (γmean), near-maximum gamma (γ1%) and the difference in isocenter dose (ΔDisoc) as metrics.
Results: Out of a total of 42 serious errors, the number of errors detected was 33 (79%), and 27 out of 30 (90%) if setup errors are not included. The system was able to pick up errors of 5 mm movement of a leaf bank, a wrong collimator rotation angle and a wrong photon beam energy. A change in phantom thickness of 1 cm was detected for all cases, while only for the head-and-neck plans a 2 cm horizontal and vertical shift of the phantom were alerted. A single leaf error of 5 mm could be detected for the lung plan only.
Conclusion: Although performed for a limited number of cases and error types, this study shows that EPID-based 3D transit dosimetry is able to detect a number of serious errors in dose delivery, leaf bank position and patient thickness during VMAT delivery. Errors in patient setup and single leaf position can only be detected in specific cases.
Keywords: EPID dosimetry; Error detection; Quality assurance; VMAT.
Copyright © 2018 Associazione Italiana di Fisica Medica. Published by Elsevier Ltd. All rights reserved.
Similar articles
-
Site-specific alert criteria to detect patient-related errors with 3D EPID transit dosimetry.Med Phys. 2019 Jan;46(1):45-55. doi: 10.1002/mp.13265. Epub 2018 Nov 16. Med Phys. 2019. PMID: 30372521
-
EPID sensitivity to delivery errors for pre-treatment verification of lung SBRT VMAT plans.Phys Med. 2019 Mar;59:37-46. doi: 10.1016/j.ejmp.2019.02.007. Epub 2019 Feb 26. Phys Med. 2019. PMID: 30928064
-
A novel approach to SBRT patient quality assurance using EPID-based real-time transit dosimetry : A step to QA with in vivo EPID dosimetry.Strahlenther Onkol. 2020 Feb;196(2):182-192. doi: 10.1007/s00066-019-01549-z. Epub 2020 Jan 10. Strahlenther Onkol. 2020. PMID: 31925465
-
A Swiss cheese error detection method for real-time EPID-based quality assurance and error prevention.Med Phys. 2017 Apr;44(4):1212-1223. doi: 10.1002/mp.12142. Epub 2017 Mar 17. Med Phys. 2017. PMID: 28134989
-
AAPM Task Group Report 307: Use of EPIDs for Patient-Specific IMRT and VMAT QA.Med Phys. 2023 Aug;50(8):e865-e903. doi: 10.1002/mp.16536. Epub 2023 Jun 29. Med Phys. 2023. PMID: 37384416 Free PMC article. Review.
Cited by
-
In vivo dosimetry in external beam photon radiotherapy: Requirements and future directions for research, development, and clinical practice.Phys Imaging Radiat Oncol. 2020 Aug 29;15:108-116. doi: 10.1016/j.phro.2020.08.003. eCollection 2020 Jul. Phys Imaging Radiat Oncol. 2020. PMID: 33458335 Free PMC article. Review.
-
Assessing the impact of adaptations to the clinical workflow in radiotherapy using transit in vivo dosimetry.Phys Imaging Radiat Oncol. 2023 Jan 30;25:100420. doi: 10.1016/j.phro.2023.100420. eCollection 2023 Jan. Phys Imaging Radiat Oncol. 2023. PMID: 36820237 Free PMC article.
-
In vivo EPID-based daily treatment error identification for volumetric-modulated arc therapy in head and neck cancers with a hierarchical convolutional neural network: a feasibility study.Phys Eng Sci Med. 2024 Sep;47(3):907-917. doi: 10.1007/s13246-024-01414-z. Epub 2024 Apr 22. Phys Eng Sci Med. 2024. PMID: 38647634
-
Evaluation of interfraction setup variations for postmastectomy radiation therapy using EPID-based in vivo dosimetry.J Appl Clin Med Phys. 2019 Oct;20(10):43-52. doi: 10.1002/acm2.12712. Epub 2019 Sep 21. J Appl Clin Med Phys. 2019. PMID: 31541537 Free PMC article.
-
Epid-based in vivo dose verification for lung stereotactic treatments delivered with multiple breath-hold segmented volumetric modulated arc therapy.J Appl Clin Med Phys. 2019 Mar;20(3):37-44. doi: 10.1002/acm2.12538. Epub 2019 Feb 20. J Appl Clin Med Phys. 2019. PMID: 30790439 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources