Coronary Stent Fracture: Clinical Evidence Vs. the Testing Paradigm
- PMID: 30341730
- DOI: 10.1007/s13239-018-00384-0
Coronary Stent Fracture: Clinical Evidence Vs. the Testing Paradigm
Abstract
Purpose: The United States' Food and Drug Administration (FDA) recommends that device manufacturers demonstrate 10 years of equivalent life duration for endovascular stents. Yet since the early 2000s clinical evidence of stent strut fracture defies the recommendations for these FDA approved devices. Stent strut fracture has been correlated with a higher incidence of adverse clinical events, such as in-stent thrombosis and in-stent restenosis.
Methods: This paper reviews the current clinical evidence, computational modelling relating to fatigue lifetimes, experimental testing of coronary stents, and the related regulatory guidance and standards.
Results: The scale of stent fracture is evident from the clinical data reviewed. In terms of model setups, either physical or computational the loadings, in particular, dictate the durability response.
Conclusions: The full scale of stent fracture is most likely under-reported and its assessment is dependent on detection time and detection resolution. Within the event of SF it is not necessarily consequential; further research is warranted to distinguish when the event negatively impacts the patient.
Keywords: Clinical data; Computational fatigue; Durability testing; Stent fracture.
Similar articles
-
Structural Mechanics Predictions Relating to Clinical Coronary Stent Fracture in a 5 Year Period in FDA MAUDE Database.Ann Biomed Eng. 2016 Feb;44(2):391-403. doi: 10.1007/s10439-015-1476-3. Epub 2015 Oct 14. Ann Biomed Eng. 2016. PMID: 26467552 Free PMC article.
-
Haemodynamics Study of Tapered Stents Intervention to Tapered Arteries.Cardiovasc Eng Technol. 2019 Dec;10(4):583-589. doi: 10.1007/s13239-019-00437-y. Epub 2019 Oct 15. Cardiovasc Eng Technol. 2019. PMID: 31617078
-
Experimental and Numerical Simulation of Biodegradable Stents with Different Strut Geometries.Cardiovasc Eng Technol. 2020 Feb;11(1):36-46. doi: 10.1007/s13239-019-00433-2. Epub 2019 Oct 29. Cardiovasc Eng Technol. 2020. PMID: 31664685
-
Review: Stent fracture in the drug-eluting stent era.Cardiovasc Revasc Med. 2016 Sep;17(6):404-11. doi: 10.1016/j.carrev.2016.06.002. Epub 2016 Jun 27. Cardiovasc Revasc Med. 2016. PMID: 27499060 Review.
-
Local Hemodynamic Forces After Stenting: Implications on Restenosis and Thrombosis.Arterioscler Thromb Vasc Biol. 2017 Dec;37(12):2231-2242. doi: 10.1161/ATVBAHA.117.309728. Epub 2017 Nov 9. Arterioscler Thromb Vasc Biol. 2017. PMID: 29122816 Review.
Cited by
-
Manufacturing, Processing, and Characterization of Self-Expanding Metallic Stents: A Comprehensive Review.Bioengineering (Basel). 2024 Sep 29;11(10):983. doi: 10.3390/bioengineering11100983. Bioengineering (Basel). 2024. PMID: 39451359 Free PMC article. Review.
-
Multiple type IV stent fractures resulting in an in-stent restenosis and coronary aneurysm formation - a dark side of "full-metal jacket" stenting.Postepy Kardiol Interwencyjnej. 2024 Jun;20(2):224-226. doi: 10.5114/aic.2024.139631. Epub 2024 May 14. Postepy Kardiol Interwencyjnej. 2024. PMID: 39022707 Free PMC article. No abstract available.
-
Evaluation of Ultrasonic Spray Method for Application of Sirolimus-Eluting Coating on Bioresorbable Vascular Scaffolds.Int J Mol Sci. 2025 Aug 7;26(15):7649. doi: 10.3390/ijms26157649. Int J Mol Sci. 2025. PMID: 40806774 Free PMC article.
Publication types
MeSH terms
LinkOut - more resources
Full Text Sources