Eliminating Daily Shifts, Tattoos, and Skin Marks: Streamlining Isocenter Localization With Treatment Plan Embedded Couch Values for External Beam Radiation Therapy
- PMID: 30355524
- DOI: 10.1016/j.prro.2018.08.011
Eliminating Daily Shifts, Tattoos, and Skin Marks: Streamlining Isocenter Localization With Treatment Plan Embedded Couch Values for External Beam Radiation Therapy
Abstract
Purpose: The Radiation Oncology Incident Learning System demonstrated that incorrect or omitted patient shifts during treatment are common near-misses or incidents. This single pediatric hospital quality improvement experience evaluated a markless isocenter localization workflow to improve safety and streamline treatment, obviating the need for daily shifts.
Methods and materials: Patients undergoing radiation therapy were simulated and treated with indexed immobilization devices. User origins were established at simulation based on a limited set of fixed couch-top references. In treatment planning, shifts from the user origin to the planned isocenter were converted to absolute couch parameters and embedded in the setup field parameters. Thus, the first fraction did not require any shifts. Before kilovoltage imaging, setup verification was often supplemented with surface-guided imaging. After image guidance and final couch adjustments, couch parameters could be reacquired and used for subsequent treatments. No skin marks were used.
Results: Over 3 years, approximately 300 patients were treated with over 5000 treatment fractions using this workflow. There were no wrong-site treatment errors. Approximately a dozen near-miss events related to the daily setup process occurred, largely on the first treatment. Root-cause analysis attributed errors to user origin misidentification, couch parameter miscalculation, incorrect immobilization device use, and immobilization device indexed at the wrong indexing position. Skin marks and tattoos were unnecessary. Continuous quality improvement added additional quality assurance checks, resulting in no near-miss incidents or adverse events in the preceding 12 months.
Conclusion: We minimized near-miss incidents by using limited simulation user origins, converting user origin-to-isocenter shifts to absolute couch parameters, and enforcing restrictive tolerance tables to limit delivery parameter changes, coupled with surface guidance and quality assurance tools. This technique can be applied across institutions, age ranges, and tumor types and with or without surface guidance. This workflow has removed a common treatment setup error and the need for skin marks.
Copyright © 2018 American Society for Radiation Oncology. Published by Elsevier Inc. All rights reserved.
Similar articles
-
Site-specific tolerance tables and indexing device to improve patient setup reproducibility.J Appl Clin Med Phys. 2015 May 8;16(3):5097. doi: 10.1120/jacmp.v16i3.5097. J Appl Clin Med Phys. 2015. PMID: 26103475 Free PMC article.
-
Using a daily monitoring system to reduce treatment position override rates in external beam radiation therapy.J Appl Clin Med Phys. 2022 Jul;23(7):e13629. doi: 10.1002/acm2.13629. Epub 2022 May 4. J Appl Clin Med Phys. 2022. PMID: 35506575 Free PMC article.
-
Technical Note: Dosimetric effects of couch position variability on treatment plan quality with an MRI-guided Co-60 radiation therapy machine.Med Phys. 2016 Aug;43(8):4514. doi: 10.1118/1.4955116. Med Phys. 2016. PMID: 27487868
-
Dynamic targeting image-guided radiotherapy.Med Dosim. 2006 Summer;31(2):113-25. doi: 10.1016/j.meddos.2005.12.014. Med Dosim. 2006. PMID: 16690452 Review.
-
Dosimetric effects caused by couch tops and immobilization devices: report of AAPM Task Group 176.Med Phys. 2014 Jun;41(6):061501. doi: 10.1118/1.4876299. Med Phys. 2014. PMID: 24877795 Review.
Cited by
-
A Potential Pitfall and Clinical Solutions in Surface-Guided Deep Inspiration Breath Hold Radiation Therapy for Left-Sided Breast Cancer.Adv Radiat Oncol. 2023 May 27;8(6):101276. doi: 10.1016/j.adro.2023.101276. eCollection 2023 Nov-Dec. Adv Radiat Oncol. 2023. PMID: 38047221 Free PMC article.
-
Survey on fan-beam computed tomography for radiotherapy: Current implementation and future perspectives of motion management and surface guidance devices.Phys Imaging Radiat Oncol. 2023 Dec 10;29:100523. doi: 10.1016/j.phro.2023.100523. eCollection 2024 Jan. Phys Imaging Radiat Oncol. 2023. PMID: 38187170 Free PMC article.
-
Comparison of the dosimetric accuracy of proton breast treatment plans delivered with SGRT and CBCT setups.J Appl Clin Med Phys. 2021 Sep;22(9):153-158. doi: 10.1002/acm2.13357. Epub 2021 Jul 20. J Appl Clin Med Phys. 2021. PMID: 34288378 Free PMC article.
-
Development of a taxonomy for characterising medical oncology-related patient safety and quality incidents: a novel approach.BMJ Open Qual. 2022 Jul;11(3):e001828. doi: 10.1136/bmjoq-2022-001828. BMJ Open Qual. 2022. PMID: 35793864 Free PMC article. No abstract available.
-
Automatic couch position calculation using eclipse scripting for external beam radiotherapy.J Appl Clin Med Phys. 2021 Feb;22(2):77-84. doi: 10.1002/acm2.13159. Epub 2021 Jan 13. J Appl Clin Med Phys. 2021. PMID: 33440075 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources