Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data
- PMID: 30357588
- DOI: 10.1007/s10633-018-9660-z
Evaluation of light- and dark-adapted ERGs using a mydriasis-free, portable system: clinical classifications and normative data
Abstract
Purpose: The full-field electroretinogram (ff-ERG) is a widely used clinical tool to evaluate generalized retinal function by recording electrical potentials generated by the cells in the retina in response to flash stimuli and requires mydriasis. The purpose of this study was to determine the intra-visit reliability and diagnostic capability of a handheld, mydriasis-free ERG, RETeval (LKC Technologies, Gaithersburg, MD, USA), in comparison with the standard clinical ff-ERG by measuring responses recommended by the International Society for Clinical Electrophysiology of Vision (ISCEV).
Methods: This prospective, cross-sectional study included 35 patients recruited at the Hospital for Sick Children (median age = 17, range 11 months-69 years) who had undergone a clinical ff-ERG according to ISCEV standards. For RETeval (n = 35), pupils were undilated in most (n = 29) and sensor strip electrodes were placed under the inferior orbital rim. Stimulus settings on RETeval were equivalent to those used in the clinical ERG. Fifty-seven control participants (median age = 22, range 8-65 years) underwent undilated RETeval ERG to establish standard values for comparison. Patient waveform components with amplitudes < 5th percentile, or implicit times > 95th percentile of normal relative to control data were classified as abnormal for the RETeval system.
Results: The RETeval system demonstrated a high degree of within-visit reliability for amplitudes (ICC = 0.82) and moderate reliability for implicit times (ICC = 0.53). Cohen's Kappa analysis revealed a substantial level of agreement between the diagnostic capability of RETeval in comparison with clinical ff-ERG (k = 0.82), with a sensitivity and specificity of 1.00 and 0.82, respectively. Pearson's correlations for clinical ERG versus RETeval demonstrated a positive correlation for amplitudes across the rod (r = 0.65) and cone (r = 0.74) ERG waveforms. Bland-Altman plots showed no bias between the mean differences across all amplitude and implicit time parameters of the two systems.
Conclusions: The present study demonstrated that RETeval is a reliable tool with reasonable accuracy in comparison with the clinical ERG. The portable nature of RETeval system enables its incorporation at resource-limited centers where the ff-ERG is not readily available. The avoidance of sedation and pupillary dilation are added advantages of RETeval ERG.
Keywords: Electroretinography; Handheld electroretinography; Mydriasis-free; RETeval.
Similar articles
-
Measuring differences in the ERG in myopia using the RETeval device with skin electrodes.Ophthalmic Physiol Opt. 2025 May;45(3):666-680. doi: 10.1111/opo.13460. Epub 2025 Feb 13. Ophthalmic Physiol Opt. 2025. PMID: 39945246 Free PMC article.
-
Comparison of the handheld RETeval ERG system with a routine ERG system in healthy adults and in paediatric patients.Eye (Lond). 2021 Aug;35(8):2180-2189. doi: 10.1038/s41433-020-01221-2. Epub 2020 Oct 19. Eye (Lond). 2021. PMID: 33077909 Free PMC article.
-
Assessing nonsedated handheld cone flicker electroretingram as a screening test in pediatric patients: comparison to sedated conventional cone flicker electroretinogram.J AAPOS. 2019 Feb;23(1):34.e1-34.e5. doi: 10.1016/j.jaapos.2018.09.009. Epub 2019 Jan 6. J AAPOS. 2019. PMID: 30625363
-
Electroretinograms.Handb Clin Neurol. 2019;160:481-493. doi: 10.1016/B978-0-444-64032-1.00032-1. Handb Clin Neurol. 2019. PMID: 31277870 Review.
-
The Use of the RETeval Portable Electroretinography Device for Low-Cost Screening: A Mini-Review.Klin Monbl Augenheilkd. 2024 Apr;241(4):533-537. doi: 10.1055/a-2237-3814. Epub 2024 Apr 23. Klin Monbl Augenheilkd. 2024. PMID: 38653305 Review. English.
Cited by
-
Advanced Retinal Imaging and Ocular Parameters of the Rhesus Macaque Eye.Transl Vis Sci Technol. 2021 May 3;10(6):7. doi: 10.1167/tvst.10.6.7. Transl Vis Sci Technol. 2021. PMID: 34111251 Free PMC article.
-
Human retinal dark adaptation tracked in vivo with the electroretinogram: insights into processes underlying recovery of cone- and rod-mediated vision.J Physiol. 2022 Nov;600(21):4603-4621. doi: 10.1113/JP283105. Epub 2022 Jun 7. J Physiol. 2022. PMID: 35612091 Free PMC article. Review.
-
Electroretinographic Findings in Fragile X, Premutation, and Controls: A Study of Biomarker Correlations.Int J Mol Sci. 2025 Jul 16;26(14):6830. doi: 10.3390/ijms26146830. Int J Mol Sci. 2025. PMID: 40725073 Free PMC article.
-
Comparing the RETeval® portable ERG device with more traditional tabletop ERG systems in normal subjects and selected retinopathies.Doc Ophthalmol. 2023 Apr;146(2):137-150. doi: 10.1007/s10633-022-09903-w. Epub 2022 Oct 23. Doc Ophthalmol. 2023. PMID: 36273363 Free PMC article.
-
Effects of recording sequence on flicker electroretinographics recorded with natural pupils corrected for pupil area.Acta Ophthalmol. 2021 Jun;99(4):411-417. doi: 10.1111/aos.14618. Epub 2020 Sep 10. Acta Ophthalmol. 2021. PMID: 32914561 Free PMC article.
References
Publication types
MeSH terms
Substances
LinkOut - more resources
Full Text Sources
Medical