Off-label MRI examination of an MR conditional pacemaker
- PMID: 30363654
- PMCID: PMC6180873
- DOI: 10.1259/bjrcr.20150219
Off-label MRI examination of an MR conditional pacemaker
Abstract
An 18-year-old patient was implanted with an MR-conditional pacemaker using a lead plug which negated the manufacturer CE approved MR conditional status of the pacing system. The patient was referred for a lumbar spine MRI. The clinical opinion was that the risk of proceeding with the MRI was acceptable given the alternatives of: (a) changing the pacing system or (b) not performing the scan. A risk assessment revealed that the additional risk was very low and easily mitigated by an established technique. Informed consent was obtained and the scan was completed without incident following a slightly modified MR-conditional pacemaker scanning protocol.
References
-
- Medtronic, Inc., 2010 Medtronic Ensura DR MRI™ Surescan™ EN1DR01: MRI procedural information for Ensura DR MRI™ Surescan™ EN1DR01 and CapSureFix MRI™ 5086MRI: Technical Manual. Minneapolis, MN: Medtronic; [Accessed 23 March 2015] Available from: http://www.medtronic-mrisafety.co.uk/wp-content/uploads/EN3DR01_ensura-t...
-