A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars
- PMID: 30390968
- DOI: 10.1016/j.joen.2018.09.001
A Micro-computed Tomographic Evaluation of the Accuracy of 3 Electronic Apex Locators in Curved Canals of Mandibular Molars
Abstract
Introduction: This study evaluated the accuracy of 3 electronic apex locators (EALs) (CanalPro [Coltene-Endo, Cuyahoga Falls, OH], Apex ID [SybronEndo, Glendora, CA], and Root ZX Mini [J Morita, Tokyo, Japan]) in curved mesial canals of extracted mandibular molars using micro-computed tomographic (micro-CT) scanning.
Methods: The root canal length and the actual working length of 58 canals were measured using the visual method and 3-dimensional micro-CT reconstructions. The measurements of the EALs at marks "APEX/0.0" and "0.5 mark" were recorded as the electronic root canal length and the electronic working length, respectively. The absolute mean values and the percentages of distribution of the electronic measurements were compared with the actual lengths of the canals. The presence of isthmuses, accessory canals, foramen position, and the distance between the apical constriction and the apical foramen were recorded from the micro-CT scans, and their correlation to the accuracy of the EALs was determined.
Results: All electronic measurements showed high agreement with their respective gold standard, except the electronic root canal length of the Apex ID (P < .05). No difference in the percentage of precise measurements (within the ±0.5 mm) was found among the devices (χ2 and z tests, P > .05). Of the anatomic parameters evaluated, the presence of a lateral foramen negatively affected the 0.0 mark of the Apex ID (Wilcoxon test, P < .05).
Conclusions: The Root ZX Mini and CanalPro were precise for both root canal length and working length determination in mesial curved canals of mandibular molars, whereas the Apex ID was accurate for the working length when using the 0.5 mark.
Keywords: Apex ID; CanalPro; Root ZX; electronic apex locator; micro–computed tomography; working length.
Copyright © 2018 American Association of Endodontists. Published by Elsevier Inc. All rights reserved.
Similar articles
-
The Use of Micro-Computed Tomography to Determine the Accuracy of 2 Electronic Apex Locators and Anatomic Variations Affecting Their Precision.J Endod. 2016 Aug;42(8):1263-7. doi: 10.1016/j.joen.2016.04.024. Epub 2016 Jun 9. J Endod. 2016. PMID: 27291503
-
Comparison of working length determination with radiographs and four electronic apex locators.Int Endod J. 2011 Jun;44(6):510-8. doi: 10.1111/j.1365-2591.2011.01855.x. Epub 2011 Feb 10. Int Endod J. 2011. PMID: 21306402
-
Ex vivo evaluation of the accuracy of two electronic apex locators during root canal length determination in primary teeth.Int Endod J. 2008 Apr;41(4):317-21. doi: 10.1111/j.1365-2591.2007.01366.x. Epub 2008 Jan 19. Int Endod J. 2008. PMID: 18217990
-
The Precision of Electronic Apex Locators in Working Length Determination: A Systematic Review and Meta-analysis of the Literature.J Endod. 2015 Nov;41(11):1818-23. doi: 10.1016/j.joen.2015.08.012. Epub 2015 Sep 26. J Endod. 2015. PMID: 26409810
-
Electronic canal length measurement.Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995 Feb;79(2):226-31. doi: 10.1016/s1079-2104(05)80288-4. Oral Surg Oral Med Oral Pathol Oral Radiol Endod. 1995. PMID: 7614187 Review.
Cited by
-
Correlation of objective image quality and working length measurements in different CBCT machines: An ex vivo study.Sci Rep. 2020 Nov 10;10(1):19414. doi: 10.1038/s41598-020-76424-4. Sci Rep. 2020. PMID: 33173072 Free PMC article.
-
Micro-computed tomography analysis of root canal morphology and thickness of crown and root of mandibular incisors in Chinese population.Clin Oral Investig. 2022 Jan;26(1):901-910. doi: 10.1007/s00784-021-04071-8. Epub 2021 Aug 2. Clin Oral Investig. 2022. PMID: 34338894
-
The influence of periapical lesions on the repeatability of two electronic apex locators in vivo.Clin Oral Investig. 2021 Sep;25(9):5239-5245. doi: 10.1007/s00784-021-03831-w. Epub 2021 Feb 13. Clin Oral Investig. 2021. PMID: 33580848
-
Evaluation of heated sodium hypochlorite's effect on the accuracy of contemporary electronic apex locators: an in vitro study.BMC Oral Health. 2025 Apr 24;25(1):634. doi: 10.1186/s12903-025-06003-4. BMC Oral Health. 2025. PMID: 40275207 Free PMC article.
-
The Effect of Long Service Life on the Accuracy of Dentaport ZX Electronic Apex Locator.Iran Endod J. 2022;17(4):195-199. doi: 10.22037/iej.v17i4.34623. Iran Endod J. 2022. PMID: 36703688 Free PMC article.
MeSH terms
LinkOut - more resources
Full Text Sources
Miscellaneous