Accuracy of an automated blood-gas analyser operated by untrained staff
- PMID: 30464
- DOI: 10.1093/bja/50.10.1031
Accuracy of an automated blood-gas analyser operated by untrained staff
Abstract
Performance of the IL 613 automatic blood-gas analyser has been assessed using a group of 100 "analysts" with no previous training or experience in the use of the instrument. Test material consisted of blood equilibrated to a known PO2 and PCO2 in a tonometer: pH estimations were carried out on thawed aliquots of a large batch of frozen serum which were then equilibrated to a known PCO2. Eighty-six per cent of analyses were within acceptable limits of error. The largest proportion of errors was in the measurement of pH. Satisfactory results were obtained in 98% of the analyses of PO2 and PCO2. Eighty-eight per cent of operators were able to use the analyser after instruction lasting less than 1 min. These results were significantly better than those obtained in a regional survey of 16 blood-gas laboratories, staffed by trained technicians.
Similar articles
-
Comparative study of automatic blood-gas analysers and their use in analysing arterial and capillary samples.Br Med J. 1979 Jan 20;1(6157):156-8. doi: 10.1136/bmj.1.6157.156. Br Med J. 1979. PMID: 33741 Free PMC article.
-
Towards blood-gas autoanalysis an evaluation of the radiometer ABL.Br J Anaesth. 1976 May;48(5):487-94. doi: 10.1093/bja/48.5.487. Br J Anaesth. 1976. PMID: 1276024
-
[Analytical reliability of the results of determining pH, pCO2 and pO2 with the ABL 300 (Radiometer) automatic pH and blood gas analyzer].Vutr Boles. 1988;27(4):77-81. Vutr Boles. 1988. PMID: 3145645 Bulgarian.
-
What is new in blood-gas analysis?Br J Anaesth. 1978 Jan;50(1):51-62. doi: 10.1093/bja/50.1.51. Br J Anaesth. 1978. PMID: 23819 Review. No abstract available.
-
Sources of error in acid-base analysis from a blood gas analyser result: a narrative review.J S Afr Vet Assoc. 2022 Nov;93(2):89-98. doi: 10.36303/JSAVA.163. Epub 2022 Jun 24. J S Afr Vet Assoc. 2022. PMID: 35934908 Review.
Cited by
-
Oxygen tension measurement using an automatic blood gas analyser.J R Soc Med. 1981 Aug;74(8):595-600. doi: 10.1177/014107688107400808. J R Soc Med. 1981. PMID: 7288796 Free PMC article.
-
Electrolyte measurement by clinicians.Intensive Care Med. 1987;13(6):411-5. doi: 10.1007/BF00257686. Intensive Care Med. 1987. PMID: 3668075
-
Clinical biochemistry nearer the patient.Br Med J (Clin Res Ed). 1983 Apr 9;286(6372):1166-7. doi: 10.1136/bmj.286.6372.1166. Br Med J (Clin Res Ed). 1983. PMID: 6404375 Free PMC article. No abstract available.
MeSH terms
Substances
LinkOut - more resources
Full Text Sources