. 2019 Oct;32(5):841-848.
doi: 10.1007/s10278-018-0157-9.

Remote Real-Time Ultrasound Supervision via Commercially Available and Low-Cost Tele-Ultrasound: a Mixed Methods Study of the Practical Feasibility and Users' Acceptability in an Emergency Department

Affiliations

Remote Real-Time Ultrasound Supervision via Commercially Available and Low-Cost Tele-Ultrasound: a Mixed Methods Study of the Practical Feasibility and Users' Acceptability in an Emergency Department

Stig Holm Jensen et al. J Digit Imaging. 2019 Oct.

Abstract

Minor emergency departments (ED) struggle to access sufficient expertise to supervise learners of lung and cardiac point-of-care ultrasound (POCUS). Using tele-ultrasound (tele-US) for remote supervision may remedy this situation. We aimed to evaluate the feasibility of real-time supervision via tele-US when applied to an everyday ED clinic. We conducted a mixed methods study that assessed practical feasibility, determined performance, and explored users' acceptability of supervision via tele-US. Technical performance was assessed quantitatively by the ratio in mean gray value between images on site and as received by the supervisor, and by after-compression frame rate. Qualitatively, 12 exploratory semi-structured interviews were conducted with exposed junior doctors and supervisors. Remote supervision via tele-US was performed with 10 junior doctors scanning 45 included patients. During performance assessment, neither alternating internet connection nor software significantly changed the mean gray value ratio. The lowest median frame rate of 4.6 (interquartile range [IQR]: 3.1-5.0) was found by using a 4G internet connection; the highest of 28.5 (IQR: 28.5-29.0) was found with alternative computer and local area network internet connection. In interviews, supervisors stressed the importance of preserving frame rate, and junior doctors emphasized a need for shared ultrasound terminology. In the qualitative analysis, setup mobility, accessibility, and time consumption were emphasized as being of key importance for future clinical implementations. Remote supervision via a commercially available and low-cost tele-US setup is operational for both junior doctors and supervisors when applied to lung and cardiac POCUS scans of hospitalized patients.

Keywords: Point-of-care ultrasound; Tele-ultrasound.

PubMed Disclaimer

Conflict of interest statement

The authors declare that there are no conflicts of interest. This research was not sponsored by any company. None of the authors have any financial relationship with Apple; Lenovo; MM Vision; Visicom Media; Logitech; GE Healthcare; Marshall; Epiphan Video; Skype; National Institutes of Health, USA; Ookla; or any of their products. The authors have full control of data.

Figures

Fig. 1
Fig. 1
Tele-ultrasound setup: (a) ultrasound system, (b) web cameras, (c) video grabber, (d) headset, (e) on-site laptop, (f) internet connection, and (g) remote laptop
Fig. 2
Fig. 2
Supervisor’s view: two-angled overview video (left) and US image (right)
Fig. 3
Fig. 3
Frames per second: Median, IQR, minimum, and maximum. ref.: equipment used in the clinical implementation, and the reference used in statistical comparison. * = p < 0.05, ** = p < 0.01 from Kruskall–Wallis one-way analysis of variance
Fig. 4
Fig. 4
Mean gray value (MGV) ratio: Median, IQR, minimum, and maximum. ref.: equipment used in the clinical implementation, and the reference used in statistical comparison. * = p < 0.05, ** = p < 0.01 from Kruskall–Wallis one-way analysis of variance

Similar articles

Cited by

References

    1. Labovitz AJ, Noble VE, Bierig M, Goldstein SA, Jones R, Kort S, Porter TR, Spencer KT, Tayal VS, Wei K. Focused Cardiac Ultrasound in the Emergent Setting: A Consensus Statement of the American Society of Echocardiography and American College of Emergency Physicians. J Am Soc Echocardiogr. 2010;23(12):1225–1230. doi: 10.1016/j.echo.2010.10.005. - DOI - PubMed
    1. International Liaison Committee on Lung Ultrasound for the International Consensus Conference on Lung U. Volpicelli G, Elbarbary M, Blaivas M, Lichtenstein DA, Mathis G, et al. International evidence-based recommendations for point-of-care lung ultrasound. Intensive Care Med. 2012;38(4):577–591. doi: 10.1007/s00134-012-2513-4. - DOI - PubMed
    1. Moore CL, Copel JA. Point-of-care ultrasonography. N Engl J Med. 2011;364(8):749–757. doi: 10.1056/NEJMra0909487. - DOI - PubMed
    1. Kirkpatrick AW, McKee I, McKee JL, Ma I, McBeth PB, Roberts DJ, et al. Remote just-in-time telementored trauma ultrasound: a double-factorial randomized controlled trial examining fluid detection and remote knobology control through an ultrasound graphic user interface display. Am J Surg. 2016;211(5):894–902 e1. doi: 10.1016/j.amjsurg.2016.01.018. - DOI - PubMed
    1. Dyer D, Cusden J, Turner C, Boyd J, Hall R, Lautner D, Hamilton DR, Shepherd L, Dunham M, Bigras A, Bigras G, McBeth P, Kirkpatrick AW. The clinical and technical evaluation of a remote telementored telesonography system during the acute resuscitation and transfer of the injured patient. J Trauma. 2008;65(6):1209–1216. doi: 10.1097/TA.0b013e3181878052. - DOI - PubMed

Publication types