In-vivo and numerical analysis of the eigenmodes produced by a multi-level Tic-Tac-Toe head transmit array for 7 Tesla MRI
- PMID: 30481187
- PMCID: PMC6258503
- DOI: 10.1371/journal.pone.0206127
In-vivo and numerical analysis of the eigenmodes produced by a multi-level Tic-Tac-Toe head transmit array for 7 Tesla MRI
Abstract
Radio-frequency (RF) field inhomogeneities and higher levels of specific absorption rate (SAR) still present great challenges in ultrahigh-field (UHF) MRI. In this study, an in-depth analysis of the eigenmodes of a 20-channel transmit Tic-Tac-Toe (TTT) RF array for 7T neuro MRI is presented. The eigenmodes were calculated for five different Z levels (along the static magnetic field direction) of the coil. Four eigenmodes were obtained for each Z level (composed of 4 excitation ports), and they were named based on the characteristics of their field distributions: quadrature, opposite-phase, anti-quadrature, and zero-phase. Corresponding finite-difference time-domain (FDTD) simulations were performed and experimental B1+ field maps were acquired using a homogeneous spherical phantom and human head (in-vivo). The quadrature mode is the most efficient and it excites the central brain regions; the opposite-phase mode excites the brain peripheral regions; anti-quadrature mode excites the head periphery; and the zero-phase mode excites cerebellum and temporal lobes. Using this RF array, up to five eigenmodes (from five different Z levels) can be simultaneously excited. The superposition of these modes has the potential to produce homogeneous excitation with full brain coverage and low levels of SAR at 7T MRI.
Conflict of interest statement
Tiejun Zhao is employed by Siemens Medical Solutions. There are no patents or products in development to declare. This does not alter our adherence to PLOS ONE policies on sharing data and materials.
Figures
References
-
- Tropp J. Image brightening in samples of high dielectric constant. J Magn Reson. 2004;167(1):12–24. 10.1016/j.jmr.2003.11.003 . - DOI - PubMed
-
- Ibrahim TS, Mitchell C, Abraham R, Schmalbrock P. In-depth study of the electromagnetics of ultrahigh-field MRI. Nmr Biomed. 2007;20(1):58–68. Epub 2006/09/29. 10.1002/nbm.1094 . - DOI - PubMed
-
- Vaughan JT, Garwood M, Collins CM, Liu W, DelaBarre L, Adriany G, et al. 7T vs. 4T: RF power, homogeneity, and signal-to-noise comparison in head images. Magn Reson Med. 2001;46(1):24–30. . - PubMed
-
- Ibrahim TS, Tang L. Insight into RF power requirements and B1 field homogeneity for human MRI via rigorous FDTD approach. J Magn Reson Imaging. 2007;25(6):1235–47. Epub 2007/05/24. 10.1002/jmri.20919 . - DOI - PubMed
-
- Röschmann P. Radiofrequency penetration and absorption in the human body: Limitations to high‐field whole‐body nuclear magnetic resonance imaging. Medical physics. 1987;14(6):922–31. 10.1118/1.595995 - DOI - PubMed