Review
. 2018 Nov 13:6:504.
doi: 10.3389/fchem.2018.00504. eCollection 2018.

Traditional Calibration Methods in Atomic Spectrometry and New Calibration Strategies for Inductively Coupled Plasma Mass Spectrometry

Affiliations
Review

Traditional Calibration Methods in Atomic Spectrometry and New Calibration Strategies for Inductively Coupled Plasma Mass Spectrometry

Jake A Carter et al. Front Chem. .

Abstract

Applications, advantages, and limitations of the traditional external standard calibration, matrix-matched calibration, internal standardization, and standard additions, as well as the non-traditional interference standard method, standard dilution analysis, multi-isotope calibration, and multispecies calibration methods are discussed.

Keywords: interference standard method; internal standardization; matrix-matched calibration; multi-isotope calibration; multispecies calibration; standard additions; standard dilution analysis.

PubMed Disclaimer

Figures

Figure 1
Figure 1
SA calibration plot using four addition points (x1-x4). Extrapolation is represented by a broken line, and x0 is the sample with no standard addition.
Figure 2
Figure 2
Calibration plot used to determine Si in a 1% v/v HNO3 matrix by ICP-QMS (m/z = 28). No collision/reaction gas was used.
Figure 3
Figure 3
Calibration plot used to determine Si in a 1% v/v HNO3 matrix by ICP-QMS (m/z = 28) using the IFS method. The 38Ar+ ion was used as IFS species, and the analyte-to-IFS signal ratio was plotted on the y-axis. No collision/reaction gas was used.
Figure 4
Figure 4
SDA plot for the determination of Ba and Cu in Oyster Tissue (NIST 1566b) by ICP-MS. Yttrium was used as internal standard, and He flowing at 3.5 mL/min was used in the collision/reaction cell.
Figure 5
Figure 5
Calibration plots created from the SDA region (negative slopes) in Figure 4. The concentrations of Ba, Cu, and Y added to S1 were 50 μg/L each.
Figure 6
Figure 6
MICal plot used to determine Cd in a tap water sample spiked with 10 μg/L of the analyte. The concentration of Cd standard added to S1 [C(A)Std] was 30 μg/L. He flowing at 3.5 mL/min was used in the collision/reaction cell.
Figure 7
Figure 7
MICal plot used to determine Cd in a tap water sample spiked with 10 μg/L of the analyte. S1 and S2 were prepared with 70% of sample and 30% of standard solution or blank. The concentration of Cd standard added to S1 [C(A)Std] was 30 μg/L. He flowing at 3.5 mL/min was used in the collision/reaction cell.
Figure 8
Figure 8
Summary of the most adequate general applications for EC, IS, SA, IFS, SDA, MICal, and MSC.

Similar articles

Cited by

References

    1. Abbyad P., Tromp J., Lam J., Salin E. (2001). Optimization of the technique of standard additions for inductively coupled plasma mass spectrometry. J. Anal. Atomic Spectrom. 16, 464–469. 10.1039/b100672j - DOI
    1. ABNT NBR 15556 (2008). Fat Oil and Derivatives. Fatty Acid Methyl Esters. Determination of the Sodium, Potassium, Calcium and Magnesium Contents by Atomic Absorption Spectrometry. São Paulo.
    1. Aguirre M. Á., Kovachev N., Almagro B., Hidalgo M., Canals A. (2010). Compensation for matrix effects on ICP-OES by on-line calibration methods using a new multi-nebulizer based on Flow Blurring® technology. J. Anal. Atomic Spectrom. 25:1724 10.1039/c004854b - DOI
    1. Alfarraj B. A., Sanghapi H. K., Bhatt C. R., Yueh F. Y., Singh J. P. (2018). Qualitative analysis of dairy and powder milk using laser-induced breakdown spectroscopy (LIBS). Appl. Spectrosc. 72, 89–101. 10.1177/0003702817733264 - DOI - PubMed
    1. Almeida J. S., Brand C., Santos G. L. (2016). Fast sequential determination of manganese and chromium in vegetable oil and biodiesel samples by high-resolution continuum source graphite. Anal. Methods 8, 3249–3254. 10.1039/c6ay00165c - DOI

LinkOut - more resources