. 2019 May;29(5):2573-2580.
doi: 10.1007/s00330-018-5800-9. Epub 2018 Nov 28.

Double-inversion recovery with synthetic magnetic resonance: a pilot study for assessing synovitis of the knee joint compared to contrast-enhanced magnetic resonance imaging

Affiliations

Double-inversion recovery with synthetic magnetic resonance: a pilot study for assessing synovitis of the knee joint compared to contrast-enhanced magnetic resonance imaging

Jisook Yi et al. Eur Radiol. 2019 May.

Abstract

Objectives: To investigate the agreement between double-inversion recovery (DIR) with synthetic magnetic resonance imaging (MRI) and T1-weighted contrast-enhanced (CE)-MRI for the assessment of knee synovitis.

Methods: T1-weighted CE-MRI and synthetic MRI of 30 patients were compared. Synthetic DIR image reconstruction was performed with two inversion times (280-330 ms and 2800-2900 ms). Subjective image quality, visibility of synovium, detection of synovitis, and total synovitis score in the knee joint were evaluated on both MR images. The relative signal intensity (SI) and relative contrast of synovium, joint effusion, and bone marrow for two imaging were assessed. Differences in data between two imaging were assessed by using Wilcoxon's signed-rank test and chi-square test/Fisher's exact test. Interobserver agreement was expressed as weighted kappa value. Accuracy of synthetic DIR image was calculated by using CE-MRI as reference standard.

Results: T1-weighted CE-MRI yielded better image quality than synthetic DIR imaging (p < 0.001). Interobserver agreements for detecting synovitis diagnosis/sum of the synovitis score were moderate to almost perfect (κ = 0.58/0.44, synthetic DIR; κ = 0.83/0.65, T1-weighted CE-MRI). There were no statistical differences in visibility of synovium (p = 0.058-0.190), detection of synovitis (p < 0.001), and relative SI of structures between two imaging (p = 0.086-0.360). Synovium-to-effusion contrast was higher in synthetic DIR (p = 0.003) and synovium-to-bone marrow contrast was higher in CE-MRI (p < 0.001).

Conclusion: Synthetic DIR imaging showed a moderate degree of interobserver agreement and good accuracy for detecting synovitis. Though it has limitations, it may play a role in imaging of degenerative joint disease or larger cohort scientific studies where gadolinium application is not feasible.

Key points: • Synthetic double-inversion recovery (DIR) imaging avoids the use of contrast agent. • There was no significant difference between T1-weighted CE-MRI and synthetic DIR imaging in evaluating presence of synovitis in knee joint. • Synthetic DIR imaging showed moderate degree of interobserver agreement and good accuracy for detecting synovitis compared to CE-MRI, and it may facilitate evaluation of some regions of peripatellar synovitis.

Keywords: Contrast media; Knee; Magnetic resonance imaging; Synovitis.

PubMed Disclaimer

Similar articles

Cited by

References

    1. Radiology. 2014 Mar;270(3):834-41 - PubMed
    1. Rheumatology (Oxford). 2005 Dec;44(12):1569-73 - PubMed
    1. AJNR Am J Neuroradiol. 2016 Jun;37(6):1023-9 - PubMed
    1. AJR Am J Roentgenol. 2009 Oct;193(4):1124-7 - PubMed
    1. Ann Rheum Dis. 2005 Sep;64(9):1263-7 - PubMed

LinkOut - more resources