Ear measurement of temperature is only useful for screening for fever in an adult emergency department
Abstract
Background: A new generation of ear thermometers with preheated tips and several measurements points should allow a more precise temperature measurement. The aim of the study was to evaluate if the ear temperature measured by this ear thermometer can be used to screen for fever and whether the thermometer is in agreement with the rectal temperature and if age, use of hearing devices or time after admission influences the temperature measurements.
Methods: Open cross-sectional clinical single site study patients, > 18 years old, who were acutely admitted to the short stay unit at the ED. A sample size of 99 patient per subgroup was recruited as random convenience series. As ear thermometer Braun Thermoscan Pro 4000® and as rectal thermometer Omron Flex Temp Smart ® was used. For different cut off of temperature the AUC was calculated and Bland-Altman analysis for calculation of 95% limits of agreement with rectal temperature, with subgroup analysis concerning age, time span from admission time and use of hearing aid.
Results: Among 599 patients the sensitivity to detect fever with an ear thermometer varied between 68 and 70% with AUC from 0.88-0.97. If the ear temperature was ≥37.5 oC, the sensitivity to detect patients with ≥38.0 oC rectally was 95% which raised to 100% for a rectal temperature of ≥38.3 oC. For the ear thermometer's ability to determine the exact temperature the 95% limits of agreement were +/- 0.8 oC. with no influence from age, duration of hospital stay or hearing aids.
Conclusion: The examined ear thermometer is able to detect fever, defined as ≥38 oC rectally in an adult ED population by using an ear cut-point of 37.5 oC, but not to measure the exact temperature. Used in this way around a fifth of the patients will still be in need of a rectal temperature measurement, but less than 5% with fever ≥38.0 oC will remain undetected and none with fever ≥38.3 oC. Age, admission time and use of hearing aid did not influence the temperature measurements.
Trial registration: Clinical Trials: ID NCT02977481 , date 11/18/2016.
Keywords: Fever; Rectal temperature; Temperature measurement; Tympanic.
Conflict of interest statement
Ethics approval and consent to participate
Approval from the Regional Committees on Health Research Ethics for Southern Denmark was waived after request (S-20160154). The study was registered with the Danish Data Protection Agency (16/39984).
Eligible patients were informed about the study and written information was provided. If the patient required time to consider participation, up to one hour was accepted. If the patient requested to discuss participation with a lay representative before a decision to participate was made, this was secured by a telephone call to the lay representative appointed by the patient if the representative was not present. When a written consent to participate was obtained the patient was included in the study.
Consent for publication
Not applicable.
Competing interests
Prof. Christian Backer Mogensen is associate editor for BMC Emergency Medicine.
Publisher’s Note
Springer Nature remains neutral with regard to jurisdictional claims in published maps and institutional affiliations.
Figures
Similar articles
-
Forehead or ear temperature measurement cannot replace rectal measurements, except for screening purposes.BMC Pediatr. 2018 Jan 26;18(1):15. doi: 10.1186/s12887-018-0994-1. BMC Pediatr. 2018. PMID: 29373961 Free PMC article.
-
NexTemp thermometer can be used interchangeably with tympanic or mercury thermometers for emergency department use.Emerg Med Australas. 2006 Jun;18(3):245-51. doi: 10.1111/j.1742-6723.2006.00847.x. Emerg Med Australas. 2006. PMID: 16712534
-
Measurement of non-invasive rectal and ear temperature in inpatients ≥ 18 years old: A cross-sectional comparative study.Z Evid Fortbild Qual Gesundhwes. 2025 Feb;192:36-41. doi: 10.1016/j.zefq.2024.11.005. Epub 2024 Dec 13. Z Evid Fortbild Qual Gesundhwes. 2025. PMID: 39674705
-
[Body temperature measurement in daily practice].Arch Pediatr. 2005 Aug;12(8):1292-300. doi: 10.1016/j.arcped.2005.01.034. Arch Pediatr. 2005. PMID: 15993044 Review. French.
-
[Measurement of body temperature in daily practice].Rev Med Interne. 1997;18(1):30-6. doi: 10.1016/s0248-8663(97)84391-3. Rev Med Interne. 1997. PMID: 9092015 Review. French.
Cited by
-
The diagnostic accuracy of digital, infrared and mercury-in-glass thermometers in measuring body temperature: a systematic review and network meta-analysis.Intern Emerg Med. 2021 Jun;16(4):1071-1083. doi: 10.1007/s11739-020-02556-0. Epub 2020 Nov 25. Intern Emerg Med. 2021. PMID: 33237494 Free PMC article.
-
COVID-19 and thermoregulation-related problems: Practical recommendations.Temperature (Austin). 2020 Aug 6;8(1):1-11. doi: 10.1080/23328940.2020.1790971. Temperature (Austin). 2020. PMID: 33553500 Free PMC article. Review.
-
Triage body temperature and its influence on patients with acute myocardial infarction.BMC Cardiovasc Disord. 2023 Aug 4;23(1):388. doi: 10.1186/s12872-023-03372-y. BMC Cardiovasc Disord. 2023. PMID: 37542240 Free PMC article.
-
Setting up a three-stage pre-endoscopy triage during the coronavirus disease 2019 pandemic: A multicenter observational study.DEN Open. 2022 Aug 8;3(1):e159. doi: 10.1002/deo2.159. eCollection 2023 Apr. DEN Open. 2022. PMID: 35959099 Free PMC article.
-
Fever lasting 48 hours as a predictive factor of ESBL-producing bacteria in non-critically ill patients with urinary tract infection.Sci Rep. 2024 May 13;14(1):10897. doi: 10.1038/s41598-024-61824-7. Sci Rep. 2024. PMID: 38740876 Free PMC article.
References
-
- Smith R, Woldike PM. Linneberg a: [methods of first-choice for clinical temperature measurement in Danish hospital wards] Ugeskr Laeger. 2008;170(22):1930–1933. - PubMed
-
- Haugan B, Langerud AK, Kalvoy H, Froslie KF, Riise E, Kapstad H. Can we trust the new generation of infrared tympanic thermometers in clinical practice? J Clin Nurs. 2013;22(5–6):698–709. - PubMed
MeSH terms
Associated data
LinkOut - more resources
Full Text Sources
Medical
Research Materials