Sustainable Antirestenosis Effect With a Low-Dose Drug-Coated Balloon: The ILLUMENATE European Randomized Clinical Trial 2-Year Results
- PMID: 30522663
- DOI: 10.1016/j.jcin.2018.08.034
Sustainable Antirestenosis Effect With a Low-Dose Drug-Coated Balloon: The ILLUMENATE European Randomized Clinical Trial 2-Year Results
Abstract
Objectives: The aim of this study was to assess the safety and effectiveness of a next-generation low-dose drug-coated balloon (DCB) designed to optimize the amount of drug transferred into the vessel wall and to maximize the amount of time the drug resides in the vessel wall.
Background: Several randomized controlled studies evaluating various DCBs have demonstrated a significantly higher patency rate compared with noncoated percutaneous transluminal angioplasty balloons at 1 year. However, the data are limited and vary by DCB at longer follow-up time points. An earlier generation low-dose DCB failed to demonstrate significant treatment effect at 2 years, raising questions regarding the durability of low-dose DCBs.
Methods: In this prospective, multicenter trial, 294 patients were randomized (3:1) to treatment with a DCB or an uncoated percutaneous transluminal angioplasty balloon. Assessments at 2 years included primary patency with duplex ultrasonography, clinically driven target lesion revascularization, and functional outcomes.
Results: Primary patency at 2 years was significantly higher in the DCB cohort (75.9% vs. 61.0%; p = 0.025), and the rate of clinically driven target lesion revascularization was significantly lower (12.1% vs. 30.5%; p < 0.001). There were no major limb amputations in either group. The rates of all-cause (6.5% vs. 5.1%; p = 1.00) and cardiovascular-related (1.6% vs. 1.7%; p = 1.00) mortality were similar between groups. Functional improvements over baseline were sustained in both groups, with 60% fewer reinterventions in the DCB group.
Conclusions: A sustained treatment effect is achievable with a low-dose DCB with an optimized coating formulation. This trial demonstrated for the first time a statistically significantly higher primary patency rate for a low-dose DCB versus PTA at 2 years. (CVI Drug Coated Balloon European Randomized Clinical Trial; NCT01858363).
Keywords: claudication; drug-coated balloon; drug-eluting balloon; peripheral artery disease; superficial femoral artery.