Spatial shifts in frame-based Gamma Knife radiosurgery: A case for cone beam CT imaging as quality assurance using the Gamma Knife® Icon™
- PMID: 30538892
- PMCID: PMC6255723
Spatial shifts in frame-based Gamma Knife radiosurgery: A case for cone beam CT imaging as quality assurance using the Gamma Knife® Icon™
Abstract
Background: Cone beam CT (CBCT) imaging has been integrated into the most recent version of the Leksell Gamma Knife for the primary purpose to facilitate fractionated therapy.
Case description: This case study presents three patients where the CBCT system of the Gamma Knife Icon discovered potentially clinically significant frame shifts. In each case, patients were imaged with volumetric MR prior to stereotactic frame placement. Immediately following frame placement, diagnostic stereotactic CT imaging was acquired with a stereotactic indicator box attached to the frame. Following treatment planning and immediately before radiosurgery, a CBCT was acquired using the on-board imaging functionality of the Gamma Knife Icon, which provides a registration of the patient's anatomy to stereotactic space independent of that provided by the stereotactic frame/fiducials. Co-registration of the CT and CBCT provides an estimate of the difference between these two estimates of stereotactic coordinates. The vector magnitudes of the differences measured at the center of stereotactic space were 0.93mm, 2.64mm and 2.18 mm for Case 1, Case 2 and Case 3 respectively.
Conclusions: Use of the CBCT functionality of the Gamma Knife Icon to verify the consistency of frame placement can prevent clinically significant targeting errors due to frame slippage or frame adapter mounting errors, and allows any required adjustments to be made without interrupting the overall treatment workflow.
Keywords: Gamma Knife; cone beam CT; frame shift; quality assurance; radiosurgery; stereotactic frame.
Conflict of interest statement
Authors’ disclosure of potential conflicts of interest Dr. Schlesinger reports grant from Elekta Instrument, AB, outside the submitted work. Dr. Trifiletti reports other support from Novocure, outside the submitted work. Drs. Dutta, Larner, Peach and Sheehan have nothing to disclose.
Figures
References
-
- Wu A. Physics and dosimetry of the gamma knife. Neurosurg Clin N Am 1992;3(1):35-50 - PubMed
-
- Leksell L, Jernberg B. Stereotaxis and tomography. A technical note. Acta Neurochir (Wien) 1980;52(1-2):1-7 - PubMed
-
- Lindquist C, Paddick I. The Leksell Gamma Knife Perfexion and comparisons with its predecessors. Neurosurgery 2007;61(3 Suppl):130-140; discussion 140-131 - PubMed
-
- Zeverino M, Jaccard M, Patin D, Ryckx N, Marguet M, Tuleasca C, Schiappacasse L, Bourhis J, Levivier M, Bochud FO, Moeckli R. Commissioning of the Leksell Gamma Knife(R) Icon. Med Phys 2017;44(2):355-363 - PubMed
LinkOut - more resources
Full Text Sources