. 2017 Jul 14;1(4):E141-E146.
doi: 10.1055/s-0043-113999. eCollection 2017 Jul.

Intra- and Inter-Day Reliability of Body Composition Assessed by a Commercial Multifrequency Bioelectrical Impedance Meter

Affiliations

Intra- and Inter-Day Reliability of Body Composition Assessed by a Commercial Multifrequency Bioelectrical Impedance Meter

Laurent Bosquet et al. Sports Med Int Open. .

Abstract

The purpose of this study was to examine the intra- and inter-day reliability of body composition measurements provided by a commercial multifrequency bioelectrical impedance meter. Eighteen healthy, well-trained students in physical education from the same ethnic group were assessed on four consecutive days, both in the morning and in the evening. Indexes provided by the device were gathered in four categories: tissular, metabolic, hydric and ionic blocks. There was no systematic bias between repeated measures, regardless of time of day. Relative reliability was high to very high in the morning (0.72<ICC<0.99) and moderate to very high in the afternoon (0.61<ICC<0.99). Absolute reliability varied substantially between indexes (1.5%<SEM<15%). The minimum difference considered as real was proportionally altered, since it ranged from 4.2 to 41.5%. In conclusion, body composition assessed with a multifrequency bioelectrical impedance meter requires a highly standardized protocol and adjusting the cut-off value for each parameter to ascertain during athlete follow-up that a real change has occurred. This assessment should preferably be scheduled in the morning in order to decrease these cut-off values.

Keywords: absolute reliability; fat mass; fat-free mass; hydration; relative reliability.

PubMed Disclaimer

Conflict of interest statement

Conflicts of Interest The authors have no conflicts of interest that are directly relevant to the content of this manuscript.

References

    1. Atkinson G, Nevill A M. Statistical methods for assessing measurement error (reliability) in variables relevant to sports medicine. Sports Med. 1998;26:217–238. - PubMed
    1. Australian Sports Commission . Champaign (ILL): Human Kinetics; 2000. Physiological Tests for Elite Athletes.
    1. Bosquet L, Maquet D, Forthomme B, Nowak N, Lehance C, Croisier J L. Effect of the lengthening of the protocol on the reliability of muscle fatigue indicators. Int J Sports Med. 2010;31:82–88. - PubMed
    1. Currier D. Baltimore: Williams and Wilkins; 1990. Elements of Research in Physical Therapy.
    1. Deurenberg P, Yap M, van Staveren W A. Body mass index and percent body fat: A meta analysis among different ethnic groups. Int J Obes Relat Metab Disord. 1998;22:1164–1171. - PubMed